HO-480

패널 장착형 ORP 미터 (4-Wire Transmission)

산업용 ORP meter는 모든 종류의 용매의 산화-환원 전위를 측정하기 위한 ORP 변환기입니다.
무연 센서와 결합하여 환경 친화적인 측정이 가능합니다.

- 측정 범위 : 0 ~ ±2000 mV

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.
  • Includes a sensitivity correction function
    It is possible to compensate for changes in the sensitivity of the ORP electrodes.
  • A line-up of lead-free ORP electrodes
    Not only the glass body, but almost all sections of the ORP electrodes, including lead wires and soldered junctions are lead free.
  • Uses lead-free solder
    To reduce the burden on the environment, lead-free solder has been used to mount the components on the printed circuit board.
Product name ORP meter
Model HO-480
Measuring range ±2000 mV: resolution of 1 mV
Repeatability Within ±5 mV (method by reference voltage input)
Linearity Within ±5 mV (method by reference voltage input)
Transmission
output
4 mA to 20 mA DC: input/output isolated type
Maximum load resistance: 900 Ω
Transmission output range: Free setting within the measuring range
Contact output Outputs: 2 points
Alarm contact output (R1 and R2)
  Contact type: relay contact, SPDT (1c)
  Contact rating: 240 V AC, 3 A and 30 V DC, 3 A (resistance load)
  Contact function: selectable from upper/lower limit operation (ON/OFF control), alarm, and under maintenance.
Transmission
output hold feature

Selectable function

  • Holding at the latest value
  • Holding at the preset value
  • Real data output without holding
    (In the maintenance mode, the latest value is held.)
Self-diagnosis
function
  • Outside of the measuring range
  • Converter error
Ambient
temperature
−5℃ to 45℃
Relative humidity 20% to 85% (without dew condensation)
Storage
temperature
−25℃ to 65℃
Power supply Rated voltage 100 V AC to 240 V AC, 50/60 Hz, 10 VA (max.)
Structure Indoor-use panel installation type
  Panel case: ABS, Terminal: PBT
Panel: IP65 dust and water proof structure
Protective structure Panel: IP65 (IEC60529, JIS C0920)
Rear case: IP20
Teminal:IP00
Class II device (IEC61010-1)
Pollution level 2 (IEC61010-1)
Conforming
standards
CE Marking EMC Directives EN61326-1
Low Voltage Directive EN61010-1
RoHS Directive: EN50581
Class A, Industrial electromagnetic environment
  Immunity (Industrial electromagnetic environment)
  Noise increase ±0.2 pH or less
FCC Rule FCC Part15
External
dimensions
48 (W) mm × 96 (H) mm × 115 (D) mm
Case depth: approx. 105 mm
Mass Approx. 400 g

제품 문의

HORIBA제품의 자세한 정보를 원하시면, 아래의 양식에 내용을 입력을 부탁드립니다.

* 는 필수입력항목입니다.

Corporate