| Product name | Dissolved oxygen converter |
|---|
| Model | HD-200 |
|---|
| Combined sensor | 5505, 5510 (A bipolar diaphragm type polarography sensor) |
|---|
| Combined probe | DO-1100 |
|---|
Measurement range | Dissolved oxygen concentration | 0 mg/L to 20 mg/L (Display range: 0 mg/L to 22 mg/L) |
|---|
| Saturation degree | 0% to 200% (Display range: 0% to 200%) |
| Temperature | 0℃ to 50℃ (Display range: -10℃ to 110℃) |
Display resolution | Dissolved oxygen concentration | 0.01 mg/L |
|---|
| Saturation degree | 1% |
| Temperature | 0.1℃ |
| Performance | Dissolved oxygen concentration | Repeatability | Within ±0.5% of the full scale (response for equivalent input) |
|---|
| Linearity | Within ±0.5% of the full scale (response for equivalent input) |
| Temperature | Repeatability | ±0.5℃(response for equivalent input) |
| Linearity | ±0.5℃(response for equivalent input) |
Transmission output | Number of output | 2 (The negative terminals for transmission outputs are internally connected at the same electric potential) |
|---|
| Output type | 4 mA to 20 mA DC: input/output isolated type |
| Load resistance | Maximum: 900Ω |
| Linearity | Within ±0.08 mA (output only) |
| Repeatability | Within ±0.02 mA (output only) |
| Output range | Output 1 | Dissolved oxygen concentration: Selection from preset ranges or free range input within measuring range. |
| Output 2 | Free setting within a range between -20℃ and 130℃ |
| Occasional out for error | Hold or burnout to either 3.8 mA or 21 mA |
| Transmission hold | In the maintenance mode, transmission signal is held at the latest value or preset value. In the calibration mode, transmission signal can be alive or held. |
| Contact output | Number of output | 3 |
|---|
| Output type | No-voltage contact output |
| Contact type | Relay contact; SPDT (1c) |
| Output capacity | 250 V AC 3 A, 30 V DC 3 A (resistance load) |
Contact function | R1 and R2 | Selectable from upper limit alarm, lower limit alarm, ON/OFF control, currently holding of transmission output, and cleaning output (The contact is closed during alarm operation, opened normally and while the power is down). |
| FAIL | Error alarm (Closed in the normal state, opened in the failure state or while the power is down.) |
| Alarm setting range | Selectable from dissolved oxygen concentration, and temperature |
- Setting range: Within the measurement range
- Delay time: 0 to 600 seconds
|
| Control setting range | - Setting range: 0 mg/L to 20 mg/L
- Control width: 0.02 mg/L to 200 mg/L(±0.01 to ±0.50 mg/L)
|
Cleaning output | Number of output points | 1 |
|---|
| Output type | Contact output with voltage (output of connected power supply voltage) |
| Contact type | Relay contact; SPST (1a) |
| Output contact capacity | 250 V AC 0.5 A |
| External instrument | Solenoid valve for cleaning control |
| Settings | Cleaning period | 0.1 hours to 168.0 hours |
| Cleaning time | 2 seconds to 600 seconds |
| Hold time | 2 seconds to 600 seconds |
| Timer accuracy | Within 2 minutes per month |
Description of cleaning operation | One of the following operations. - Operation of internal timer only
- Operation of internal timer and external input command
- Enable the internal timer only when external input is active.
- The cleaning trigger signal (The external trigger signal must be longer than 2 seconds)
|
| Contact input | Number of input | 1 |
|---|
| Contact type | No-voltage “a” contact of open collector |
| Conditions | ON resistance: 100 Ω max. Open voltage: 24 V DC Short-circuit current: 12 mA DC max. |
| Contact function | External input for cleaning or transmission holding if cleaner is not attached. |
Transmission capability | Communication type | RS-485 |
|---|
| Signal type | 2 wire system, isolated from the input circuit Not isolated from transmission circuit |
Temperature compensation | Applicable temperature element | Platinum resistor: 1 kΩ (0℃) (The temperature sensor is built into the probe DO-1000.) |
|---|
Temperature compensation range | 0℃ to 50℃ |
| Temperature calibration | 1 point calibration comparing to reference thermometer |
| Calibration | Calibration method | Span calibration in the air or DO saturated aquaous solution |
|---|
| Number of calibration points | Atmospheric calibration: 1 point (Zero electric calibration is carried out automatically) Saturation liquid calibration : 2 points (zero calibration is omitted) |
| Additional capabilities | Salinity compensation (0.0% to 5.0%) Automatic detection of calibration failure (Zero error, Sensitivity error) Calibration history (Elapsed days from the last calibration which ever zero or span, zero shift, span sensitivity) |
| Self-check | Calibration error | Zero calibration error, sensitivity error, outside of temperature calibration range |
|---|
| Electrode diagnostic error | Sensor error (damage to diaphragm), disconnection of sensor (or damage to sealing), temperature sensor short-circuit, and electric discontinuity of temperature sensor |
| meter error | CPU error, CPU error, ADC error, memory error |
| Operating temperature range | -20℃ to 55℃ (without freeze) |
|---|
| Operating humidity range | Relative humidity: 5% to 90% (without condensation) |
|---|
| Storage temperature | -25℃ to 65℃ |
|---|
| Power supply | Rated power supply voltage | 100 V to 240 V AC ±10% 50/60 Hz |
|---|
| Power consumption | 15 VA (max) |
| Others | With power switch for maintenance use |
Compatible standards | CE marking | EMC :EN61326-1 Class A, Industrial electromagnetic environment Safety :EN61010-1 RoHS :EN50581 9. Industrial monitoring and control instruments |
|---|
| FCC rules | Part15 Class A |
| Structure | Installation | Outdoor installation type |
|---|
| Installation method | Mounted on 50A pole or wall |
| Protection code | IP65 | IEC60529, JIS C0920 |
| Case material | Aluminum alloy (coated with epoxy-denatured melamine resin) |
| Material of fittings | SUS304 |
| Material of hood | SUS304 stainless steel (coated with epoxy-denatured melamine resin) |
| Material of window | Polycarbonate |
| Display element | Reflective monochrome LCD |
| External dimensions | | 180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets) |
|---|
| Mass | | Main body: Approx. 3.5 kg; cover and mounting brackets: Approx. 1 kg |
|---|