| Product name | Industrial pH meter |
|---|
| Model | HP-480TP |
|---|
| Measuring range | 0 pH to 14 pH: 0.01 pH resolution. Temperature 0℃ to 100℃: 1℃ resolution (selectable display). |
|---|
Transmission output | 4 mA to 20 mA DC, input/output isolated type Maximum load resistance 900 Ω |
|---|
Transmission output range | Free |
|---|
Repeatability and linearity | ±0.05 pH, ±0.08 mA (for transmission output range of 0 pH to 14 pH) |
|---|
| Contact output | Outputs: 2 points Alarm contact output (R1 and R2) Contact type: relay contact, SPDT (1c) Contact rating: 240 V AC, 3 A and 30 V DC, 3 A (resistance load) Contact function: selectable from upper/lower limit operation (ON/OFF control), alarm, and maintenance. |
|---|
| Control operations | ON/OFF control - Upper and lower limits setting range: 0.00 pH to 14.00 pH
- Control width: 0.00 pH to 4.00 pH (±0.00 pH to ±2.00 pH)
Time-division proportional control - Upper and lower limits setting range: 0.00 pH to 14.00 pH
- Proportional band: 0.00 pH to 4.00 pH
- Cycle time: 5 s to 300 s
- Control output shift capability: 0% to 50% of shift volume for the cycle time
- Self-extending cycle time capability: The cycle time is extended automatically when the deviation value enters a set range (F zone) in proportion to the deviation value (This feature has no effect when the shift function has been enabled).
- F zone: 1% to 100% of the proportional band (Self-extension of the cycle time starts working when the deviation enters the above range).
- Upper limit for extending the cycle time: 0 s to 300 s.
- Maximum control volume: 50% to 100% (To be applied regardless of whether the measured value is in the proportional band or not).
|
|---|
| Calibration function | 2 points automatic calibration or manual calibration. 2 points automatic calibration: with automatic validation of the electrode characteristics. Types of standard solution: pH 2, pH 4, pH 7, pH 9 and pH 10 (JIS) Combination of standard solutions: pH 7 and one of the others Manual calibration: Freely selectable, but the difference should be over 2 pH. Temperature calibration (one point). |
|---|
Transmission output hold feature | - Previous value hold
- Optional value hold
- Continuous
Selectable from the above (However, only the previous value hold is available in the maintenance mode).
|
|---|
Self-diagnosis function | - Calibration function
Asymmetry potential error, sensitivity error, response speed error and standard solution error. - Electrode self-diagnosis
Temperature sensor short-circuit and temperature sensor disconnection. - Outside of the measuring range.
- Converter error.
|
|---|
Temperature compensation element | Selectable from compensation ON (350Ω (25℃), 500Ω (25℃),1 kΩ (0℃),6.8 kΩ (25℃) and 10 kΩ (25℃)) and compensation OFF. |
|---|
Temperature compensated range | 0℃ to 100℃ |
|---|
Ambient temperature | −5℃ to 45℃ |
|---|
| Relative humidity | 20% to 85% (without dew condensation) |
|---|
Storage temperature | −25℃ to 65℃ |
|---|
| Power supply | Rated voltage 100 V to 240 V AC, 50Hz/60 Hz, 10 VA (max.) |
|---|
| Structure | Indoor-use panel installation type Panel case: ABS, Terminal: PBT Panel: water proof structure |
|---|
| Protective structure | Panel: IP65 (IEC60529, JIS C0920) Rear case: IP20, Terminal: IP00 Class II device (IEC61010-1) Pollution level 2 (IEC61010-1) |
|---|
Conforming standards | CE Marking | EMC Directives: EN61326-1 Low Voltage Directive: EN61010-1 RoHS Directive: EN50581 | Immunity (Industrial electromagnetic environment) Noise increase ±0.2 pH or less |
|---|
| FCC Rule | FCC Part15 |
External dimensions | 48 (W) mm × 96 (H) mm × 115 (D) mm Case depth: approx. 105 mm |
|---|
| Mass | Approx. 400 g |
|---|