Select an instrument based on:
- A system that will allow the largest entrance slit width for the bandpass required.
- The highest dispersion.
- The largest optics affordable.
- Longest focal length affordable.
- Highest groove density that will accommodate the spectral range.
- Optics and coatings appropriate for specific spectral range.
- Entrance optics which will optimize etendue.
- If the instrument is to be used at a single wavelength in a non-scanning mode, then it must be possible to adjust the exit slit to match the size of the entrance slit image.
Note: f/value is not always the controlling factor of throughput. For example, light may be collected from a source at f/1 and projected onto the entrance slit of an f/6 monochromator so that the entire image is contained within the slit. Then the system will operate on the basis of the photon collection in the f/l cone and not the f/6 cone of the monochromator.