HORIBA Scientific and Analytical Instruments
Websites
Corporate Website
  • Corporate
Market Websites
  • Mobility
  • Energy and Environment Solutions
  • Life Science
  • Healthcare
  • Materials
  • Semiconductor
Countries/Regions
Americas
  • Brazil / Brasil
  • United States / United States
Asia
  • China / 中国
  • India / India
  • Japan / 日本
  • Korea / 한국
  • Singapore / Singapore
  • Taiwan / 台灣
  • Thailand / ประเทศไทย
  • Vietnam / Việt Nam
Europe
  • Austria / Österreich
  • Czech Republic / Česko
  • France / France
  • Germany / Deutschland
  • Poland / Polska
  • Russia / Россия
  • Spain / España
  • Türkiye / Türkiye
  • United Kingdom / United Kingdom
Global
  • International
Search
HORIBA Scientific and Analytical Instruments

Contact us

Find us

  • Static Light Scattering (SLS) / Laser Diffraction Particle Size Distribution Analysis
  • Dry Powder Laser Diffraction
  • Refractive Index Selection
  • Optimizing Concentration Range
  • Optimizing Ultrasonic Dispersion
  • System Verification
  • Gauge Repeatability and Reproducibility
  • Method Development Top 10 Lists
  • Bibliography
  • Related products
  1. Home
  2. Scientific and Analytical Instruments
  3. Technologies
  4. Method Development Top 10 Lists

Back to Particle Characterization Products Overview
Back to Particle Webinar List
Back to Particle Applications

Method Development Top 10 Lists

Top 10 Wet Method Development Tips

  1. Design every method for maximum precision: repeatability and reproducibility.
  2. Study different sampling techniques to help minimize error. This is particularly important for materials with small outlier populations. 
  3. Choose a solvent which has negligible reactivity with the material, wets the material without requiring copious surfactant, has a viscosity which can be pumped in the analyzer, and has a refractive index different than the material. 
  4. If surfactant is required to help wet a powder, use as little as possible to minimize bubbling. If bubbles form, use low power internal ultrasonics to remove them during measurement.
  5. Too much material has been added to the analyzer when a new sub-100 nm population appears in the particle size distribution. This is caused by multiple scattering and can be avoided by adding less material (higher %Transmittance).  
  6. Choose a refractive index by studying references (i.e. CRC Handbook), applying known information about the material, or, if no information is available, by minimizing the R-parameter.  
  7. Refractive index (RI) affects the accuracy of the particle size distribution, but has a much smaller effect on the analyzer's ability to detect changes to that distribution. In practice, even an incorrect RI can be used to successfully monitor the quality of a material or process.
  8. Study the effect of ultrasound on the material. The real-time PSD display indicates when the material has been fully dispersed by ultrasound.
  9. Evaluate the method's robustness by calculating the coefficient of variation for the D10, D50, and D90 values across a minimum of three measurement results. The LA-960 automatically makes these calculations.  
  10. Use the LA-960 Method Expert to automate and guide this important process.

Top 10 Dry Method Development Tips

  1. Design every method for maximum precision: repeatability and reproducibility.
  2. Study different sampling techniques to help minimize error. This is particularly important for materials with small outlier populations. 
  3. Dry measurements fall into two approaches: measurement everything on the chute or acquire multiple measurements from the same sampling. The former must be done when the material has a sampling bias. The latter may be used when the material does not exhibit a sampling bias. Know which approach is required for your material.
  4. Always use feedback control on the LA-960 PowderJet.
  5. Increase dispersion energy by using a smaller nozzle for the PowderJet.
  6. Choose a refractive index by studying references (i.e. CRC Handbook), applying known information about the material, or, if no information is available, by minimizing the R-parameter.  
  7. Refractive index (RI) affects the accuracy of the particle size distribution, but has a much smaller effect on the analyzer's ability to detect changes to that distribution. In practice, even an incorrect RI can be used to successfully monitor the quality of a material or process.
  8. Study the effect of air pressure on the material. Pay close attention to the coarsest and finest parts of the distribution to evaluate the extent of de-agglomeration and onset of particle breaking. 
  9. Evaluate the method's robustness by calculating the coefficient of variation for the D10, D50, and D90 values across a minimum of three measurement results. The LA-960 automatically makes these calculations.  
  10. Evaluate the method's suitability by confirming that "good" and "bad" lots of material show a difference in particle size distribution.

Browse Products

Partica

Partica

Laser Diffraction and Dynamic Imaging Particle Size and Shape Analyzer

Show More
Partica LA-960V2

Partica LA-960V2

Laser Scattering Particle Size Distribution Analyzer

Show More

Request for Information

Do you have any questions or requests? Use this form to contact our specialists.

* These fields are mandatory.

Products

  • All Products (A-Z)
  • Mobility
  • Energy and Environment
  • Life Science
  • Healthcare
  • Materials
  • Semiconductor

Applications

  • Drinking Water Utilities
  • Automotive Manufacturing
  • Semiconductor Manufacturing Process
  • Research and Testing Laboratories

Technology

  • Glow Discharge Spectroscopy
  • Pressure-based Flowmetry
  • Quadrupole Mass Spectrometry
  • Raman Spectroscopy

Service

  • On-Site Support
  • Spare Parts and Consumables

Company

  • News
  • Events
  • Career
  • History
  • Corporate Culture

Contact

  • Career Contact
  • Contact Form
  • Worldwide Locations
  • Investor Relations Contact

Terms and Conditions Privacy Notice Cookies HORIBA Group Social Media

© Copyright 2026 - All Rights Reserved