Principles
The LMNE detection principle is based on the Double Hydrodynamic Sequential System "DHSS" flow cytometry. Two characteristics for each cell passing through this system are determined: volume and absorbency.
As we can see in the Fig.1 hereunder, a part of the light is absorbed by the cell content, another part is deviated by the cell structure (scattered light), and the rest passes through or around the cell. It is the light passing through the cells that is measured (optical transmission). The absorbency is determined from this optical transmission.