discontinued 您正在瀏覽的產品已停產,但您仍可以繼續訪問此頁面以獲取相關售後服務支持。

ABX Micros 60

ABX Micros 60 HORIBA hematology analyzer video

Hematology Analyzer

Process efficiency in Hematology

  • Impedance technology, photometry
  • Microsampling: Only 10 µL
  • Smart card based memory storage (optional)
  • No compressor (stepper motor technology)

 

*This product may not be available in your country or region at this time. Please contact your HORIBA sales representative or distributor for more information. 

事業部: Healthcare
產品分類: Hematology
製造商: HORIBA ABX SAS
Base product: Micros Range

• Measurement methods: Impedance technology, photometry.

• Liquid valves (without pinching): Reliability and increased precision

• No compressor (stepper motor): Reliability, less maintenance and noise

• 8 or 18 fully automated parameters

• CBC / 3 Part DIFF

• Open or Closed Tube

Technical

Weight

31 lbs

Power supply

90/240 V - 50/60 Hz - 230 VA.

Dimensions

16.5 in x 14 x 12.5 in (HxWxD).

Printer

Dot matrix.

Throughout

60 samples/hour (on open tube)
55 samples/hour (closed tube)

Data management

Automated smart cards used for QC and patient data storage (optional).
RS 232 output.
Mono-directional connection.

Memory

Last result (with curves) or 78 results (optional smart cards based memory)

Screen

2 lines of 40 alphanumerical characters.

Keyboard

Barcode reader (optional).

 

Samples

60 samples/hour (open tube)

Parameters (16+2)

CBC Mode (18 parameters): WBC, RBC, HGB, HCT, MCV,*RDW, MCH, MCHC, PLT, MPV, PCT, *PDW, % and # of LYM, MON and GRA

Graphics

LMG histogram, RBC, PLT histograms

Sample-Volume

CBC mode: 10 µL

Stability

24 hours (CBC)

Operating sample mode

Open tube mode with 4 position tube holder (QC, 125 or 500 µL, 3 or 5 mL)

Identification

Patient I.D., Sequence#

Flags

Pathological flags (WBC, RBC, and PLT) normal limits and counting failure flags

 * RUO (Research Use Only)

Results

Parameters

Precision (cv %)

Linearity

WBC

< 2,5%

0-100 x 10³/mm3

RBC

< 2%

0-8 x 106/mm3

HGB

< 1.5%

0-26 g/dl

HCT

< 2%

0-80%

PLT (whole blood)

< 5%

0-2200 x 10³/mm3

PLT (PLT concentrate)

< 5%

0-4000 x 10³/mm3

 

Blood Controls

Reference

Description

 

2042208

ABX Minotrol 16 - 2L

2 x 2.5 mL vial

2042209

ABX Minotrol 16 - 2N

2 x 2.5 mL vial

2042202

ABX Minotrol 16 - 2H

2 x 2.5 mL vial

Calibrators

Test Reference

Description

 

2032002

ABX Minocal

2 mL vial

N: normal level, H: high level, L: low level
* Recommended by HORIBA