HR-200

現場安裝式極譜餘氯計

HORIBA 暢銷的水質計在室內應用中表現出色。

非常適合測量自來水、游泳池、水療中心和類似環境中的遊離餘氯濃度。透過結合使用清潔珠的物理拋光清潔和利用電化學反應的電化學清潔兩種方法,可以實現長期穩定的測量。

產品分類: Process Water
製造商: HORIBA Advanced Techno Co., Ltd.

Achieving stable measurements through two types of cleaning

In addition to continuous cleaning using normal abrasives (cleaning beads), electrochemical cleaning can be performed.

Measuring method Polarography
Measuring range Residual Chlorine: 0 mg/L to 3 mg/L   Resolution: 0.01 mg/L
Temperature: 0 to 50˚C   Resolution: 0.1˚C
Repeatability Residual Chlorine:Within ±0.05 mg/L or less   Temperature: Within ±0.5˚C(for equivalent input)
Linearity Residual Chlorine:Within ±0.05 mg/L or less   Temperature: Within ±0.5˚C(for equivalent input)
Transmission output Two points 4 to 20 mA DC   Input/output insulation type   Maximum load resistance 900 Ω
Contact output Three points No-voltage contact output Relay contact, SPDT
Contact capability R1, R2: Selectable from upper limit alarm, lower limit alarm, ON/OFF control, currently holding transmission output, and cleaning output.
(opened at alarm operation, closed usually, closed at power-off)
 FAIL: Error warning (normally closed; open when an error occurs; open when the power is turned OFF)
Contact input Two points Contact type: No-voltage a contact for open collector
Contact capability: Input 1: External input for transmission holding
 Input 2: Flow switch input for interlock(Open due to decreased flow)
Communication capability RS-485   Two-wire input/output insulation type (not insulated from transmission output)
Temperature compensation range 0 to 50°C
Ambient temperature 0 to 55°C
Cleaning capability Electrochemical Cleaning between Cathode and Electrochemical cleaning electrode
Calibration method Zero calibration (Zero liquid calibration)
SPAN calibration(Compare to measurement value of DPD method, Including zero electric calibration)
Self-diagnosis function Calibration error, Temperature sensor diagnostic error, Meter error
Power source 90 V to 264 V AC, 50/60 Hz 35 VA (max.) when an automatic cleaner is connected.
Construction Outdoor installation type: IP65 Protection class
Installation method: 50 A pole-mounted or wall-mounted
Material of case: Aluminum alloy (coated with epoxy modified melamine resin)
Material of mounting brackets: SUS304
Weight Approx. 4.5 kg
Equipment protected CE marking, FCC rules

留言諮詢

如您有任何疑問,请在此留下詳細需求或問題,我們將竭誠您服務。

* 這些欄位為必填項目。