尺寸
5.24
MB
〈Catalog〉Industrial Water Quality Measuring Instruments
現場安裝型電導率儀
用於測量水溶液中的電導率、CIP 的濃度調節和沖洗監測以及製程濃度計。密度測定。
適用於CIP、廢水處理、海水和水耕溶液監測的濃度調節和沖洗監測以及製程濃度計。我們的感測器系列提供適合各種樣品類型的選項,其優勢在於其堅固耐用,可承受惡劣的環境條件。
A lineup of sensors to match sample properties and applications
Two types of sensors are available: a lead type which can be directly thrown into the tank or placed in the tank with an immersion holder, and a connector type for direct insertion into piping or with a Flow-through holder.
Highly robust
Accessories required for installation and wiring possess the robustness appropriate for industrial applications, allowing them to withstand harsh environmental conditions.
Salinity concentration conversion function
The transmitter is equipped with a salinity concentration conversion function that is suitable for monitoring the salinity of seawater.
Product name | Conductivity meter | ||||
---|---|---|---|---|---|
Model | HE-200H | ||||
Sensors in combination | 4-electrode method conductivity sensor (FES Series) of cell constant 0.1/cm, or 1.0/cm | ||||
Measuring range | Measurement type | Unit | Cell constant (/cm) | 0.1 (FES-2XX, 310 series) | 1.0 (FES-125, 126) |
Conductivity | mS/cm | 0.000 to 2.000 | ○ | ○ | |
0.00 to 20.00 | ○ | ○ | |||
0.0 to 200.0 | ○ | ○ | |||
0 to 2000 | △(*Note2) | △(*Note2) | |||
Auto (*Note1) | ○ | ○ | |||
Practical region (*Note2) | 0 to 500 | 0 to 1000 | |||
Display range | 0 to 2200 | ||||
S/m | 0.000 to 0.2000 | ○ | ○ | ||
0.00 to 2.000 | ○ | ○ | |||
0.00 to 20.00 | ○ | ○ | |||
0.0 to 200.0 | △(*Note2) | △(*Note2) | |||
Auto (*Note1) | ○ | ○ | |||
Practical region (*Note2) | 0.0 to 50.0 | 0.0 to 100.0 | |||
Display range | 0.0 to 220.0 | ||||
Temperature | 0℃ to 100℃ (display range: −30℃ to 160℃) | ||||
Salinity conversion capability | Seawater | 0.00% to 4.00% | |||
NaCl | 0.0% to 20.0% | ||||
Concentration conversion capability | NaOH | 0.00% to 5.00% | |||
HNO3 | 0.00% to 5.00% | ||||
H3PO4 | 0.00% to 5.00% | ||||
Option 1 to 4 | 0.00% to 100.00% | ||||
Display resolution | Conductivity Conversion density | As shown in the measuring and conversion table. | |||
Temperature | 0.01℃ | ||||
Performance | Conductivity | Repeatability (mS/cm) | Cell constant | 0.1 (/cm) | 1.0 (/cm) |
0.00 to 20.00 | Within ±0.5% of the full scale | Within ±0.5% of the full scale | |||
20.0 to 200.0 | Within ±1.0% of the full scale | Within ±0.5% of the full scale | |||
200 to 1000 | − | Within ±1.0% of the full scale | |||
Condition | response for equivalent input | ||||
Linearity (mS/cm) | Cell constant | 0.1 (/cm) | 1.0 (/cm) | ||
0.00 to 20.00 | Within ±0.5% of the full scale | Within ±0.5% of the full scale | |||
20.0 to 200.0 | Within ±1.0% of the full scale | Within ±0.5% of the full scale | |||
200 to 1000 | − | Within ±1.0% of the full scale | |||
Condition | response for equivalent input | ||||
Performance | Temperature | Repeatability | ±0.1℃ (response for equivalent input) | ||
Linearity | ±0.5℃ (response for equivalent input) | ||||
Transmission output | Number of output | 2 (The negative terminals for transmission outputs are internally connected at the same electric potential) | |||
Output type | 4 mA to 20 mA DC: input/output isolated type | ||||
Load resistance | Maximum: 900Ω | ||||
Linearity | Within ±0.08 mA (output only) | ||||
Repeatability | Within ±0.02 mA (output only) | ||||
Output range | Output 1 | Conductivity or density converted: Free setting within a measuring range. | |||
Output 2 | Temperature: Free setting within a range between −30℃ and 160℃ | ||||
Occasional out for error | Hold or burnout to either 3.8 mA or 21 mA | ||||
Transmission hold | In the maintenance mode, transmission signal is held at the latest value or preset value. In the calibration mode, transmission signal can be alive or held. | ||||
Contact output | Number of output | 3 | |||
Output type | No-voltage contact output | ||||
Contact type | Relay contact; SPDT (1c) | ||||
Contact capacity | AC 250 V 3 A, DC 30 V 3 A | ||||
Contact function | R1 and R2 | Selectable from upper limit alarm, lower limit alarm, and currently holding of transmission output. (The contact is closed during alarm operation, opened normally and while the power is down.) | |||
FAIL | Error alarm (Closed in the normal state, opened in the failure state or while the power is down.) | ||||
Alarm setting range | Selectable from conductivity and density | ||||
| |||||
Contact input | Number of input | 2 | |||
Contact type | No-voltage "a" contact for open collector | ||||
Conditions | ON resistance: 100Ω max. Open voltage: 24 V DC Short-circuit current: 12 mA DC max. | ||||
Function of input | The input signal use can be changed. Two inputs for four output current range selection, or, one input for two current range selection and one input for holding. The display also responds for the range selection by input. | ||||
Communication function | Communication type | RS-485 | |||
Signal type | 2 wire system, isolated from the input circuit Not isolated from transmission circuit | ||||
Temperature compensation | Temperature element | Platinum resistor: 1 kΩ (0℃) | |||
Compensation method |
| ||||
Temperature compensation range | 0℃ to 100℃ (However, the compensation calculation is extended lower than 0℃ or higher than 100℃.) | ||||
Calibration | Conductivity | Based on parameter input of coefficient for the sensor cell constant. | |||
Self-check | Sensor diagnosis error | Temperature sensor short-circuit, temperature sensor disconnection, and out of the temperature measurement range | |||
Converter | CPU error, ADC error, and memory error | ||||
Operating temperature range | −20℃ to 55℃ (without freeze) | ||||
Operating humidity range | Relative humidity: 5℃ to 90℃ (without condensation) | ||||
Storage temperature | −25℃ to 65℃ | ||||
Power supply | Rated power supply voltage | 100 V to 240 V AC ±10% 50/60 Hz | |||
Power consumption | 15 VA (max) | ||||
Others | With power switch for maintenance use | ||||
Compatible standards | CE marking | EMC :EN61326-1 Class A, Industrial electromagnetic environment Safety :EN61010-1 RoHS :EN50581 9. Industrial monitoring and control instruments | |||
FCC rules | Part15 Class A | ||||
Structure | Installation | Outdoor installation type | |||
Installation method | Mount on 50 A pole or wall | ||||
Protection code | IP65 | ||||
Case material | Aluminum alloy (coated with epoxy-denatured melamine resin) | ||||
Material of fittings | SUS304 | ||||
Material of food | SUS304 stainless steel (coated with epoxy-denatured melamine resin) | ||||
Material of window | Polycarbonate | ||||
Display element | Reflective monochrome LCD | ||||
External dimensions | 180 (W) mm x 155 (H) mm x 115 (D) mm (excluding mounting brackets) | ||||
Mass | Main body: Approx. 3.5 kg; cover and mounting brackets: Approx. 1 kg |
如您有任何疑問,请在此留下詳細需求或問題,我們將竭誠您服務。