In the most fundamental sense both bandpass and resolution are used as characterization of an instrument's ability to separate adjacent spectral lines.
Assuming a continuum light source, the bandpass (BP) of an instrument is the spectral interval that may be isolated. This depends on many factors including the width of the grating, system aberrations, spatial resolution of the detector, and entrance and exit slit widths.
If a light source emits a spectrum which consists of a single monochromatic wavelength λo (Fig. 23) and is analyzed by a perfect spectrometer, the output should be identical to the spectrum of the emission (Fig. 24) which is a perfect line at precisely λo.
In reality, spectrometers are not perfect and produce an apparent spectral broadening of the purely monochromatic wavelength. The line profile now has finite width and is known as the “instrumental line profile” (instrumental bandpass) (Fig.26).
The instrumental profile may be determined in a fixed grating spectrograph configuration with the use of a reasonably monochromatic light source such as a single mode dye laser. For a given set of entrance and exit slit parameters, the grating is fixed at the proper orientation for the central wavelength of interest and the laser light source is scanned in wavelength. The output of the detector is recorded and displayed. The resultant trace will show intensity versus wavelength distribution.
For a monochromator, the same result would be achieved if a monochromatic light source is introduced into the system and the grating rotated.
The bandpass is then defined as the Full Width at Half Maximum (FWHM) of the trace, assuming monochromatic light.
Any spectral structure may be considered to be the sum of an infinity of single monochromatic lines at different wavelengths. Thus, there is a relationship between the instrumental line profile, the real spectrum and the recorded spectrum.
Let B(λ) be the real spectrum of the source to be analyzed.
Let F(λ) be the recorded spectrum through the spectrometer.
Let P(λ) be the instrumental line profile.
(34) F = B * P
The recorded function F(λ) is the convolution of the real spectrum and the instrumental line profile.
The shape of the instrumental line profile is a function of various parameters:
- width of the entrance slit
- width of the exit slit or of one pixel in the case of a
- multichannel detector
- diffraction phenomena
- aberrations
- quality of the system's components and alignment
Each of these factors may be characterized by a special function Pi(λ), each obtained by neglecting the other parameters. The overall instrumental line profile P(λ) is related to the convolution of the individual terms:
(35) P(λ) = P1(λ).P2(λ)...Pn(λ)