HORIBA uses flow cytometry and impedance measure with a reagent named ABX Fluocyte which contains a fluorescent stain the thiazole orange (TO), a cyanine stain whose excitation occurs in the visible region at λ= 488nm. It is a nucleic-acid-specific dye that enters by permeability in the cells. When it binds to RNA, there is a 3000-fold increase of fluorescence emission. It is one of dyes recommended by the NCCLS-CLSI guidelines for reticulocytes count. TO has become the most common dye employed due to its counting accuracy and practicality for clinical laboratories with commercial clinical flow cytometers with relatively low-powered visible lasers.
The degree of fluorescence emission caused by the excitation bound dye is proportional to the amount of RNA in the erythrocyte.
When the staining cell is achieved the suspension is then transferred to the laser optical bench to be measured. The laser optical bench simultaneously measures the fluorescence of the cells passing through the measuring point into the flowcell, and the volume by impedance. A cell passing through the flowcell gives 2 types of information:
- The size of the cell measured by resistivity CIS (Cell Impedance Signal)
- The Orthogonal Fluorescence Light (OFL): The fluorescence is collected using a lens focused on the optical flowcell and located at 90° from the light beam, an interferential filter specific to the TO stain selecting only the fluorescent wavelength and a photomultiplier tube.