HU-200TB-IM

Field-installation type turbidity/SS meter

For turbidity and suspended solids (SS) management in wastewater and intake water.

A turbidity meter that can be directly immersed in the tank for ease of installation. Ideal for turbidity and SS control of wastewater and intake water. Depending on the sample, the measurement method can be selected between the 90-degree transmission scattering method or the transmission method.

TURBMLSS

*

*Optional features

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Selectable measurement

Depending on the sample, the measurement method can be selected between the 90-degree transmission scattering method or the transmission method.

Compatible with a wide range of measurement

The turbidity measurement range is 0 to 4000 NTU (Formazin) and 0 to 2000 NTU (Kaolin), while the SS measurement range is 0 to 2000 mg/L (Kaolin).

Improved maintainability

Equipped with a robust built-in electric wiper cleaner that is resistant to dirt. Additionally, the sensors are equipped with a PFA resin head, making them resistant to dirt and debris. (An optional cleaning unit can also be added.)

Stable measurements

With the transmission 90-degree light scattering method, light attenuation due to light source or detector fluctuations or contamination can be canceled. Additionally, a long-life LED light source is incorporated.

A sensor that is not limited by the measurement location

Compatible with both immersion and submersion, allowing measurements to be taken in various locations without restrictions.

Converter

Product name Immersion type turbidity/SS meter
Model HU-200TB-IM Industrial-use turbidity/SS converter
Detector type SS-150 Industrial-use turbidity/SS detector
Measurement
method
Right angle scattering/
transmission
Measuring principle suitable for full range turbidity and SS within 200 mg/L by Kaorin
Transmission Measuring principle suitable for SS over 50 mg/L by Kaorin
Measurement range Turbidity Scattering/
transmission
Formazin 0 NTU to 4000 NTU
Kaolin 0 to 2000 (Display range: 2001 to 4000)
SS Scattering/
transmission
Kaolin 0 mg/L to 2000 mg/L
(Display range: 2001 mg/L to 4000 mg/L)
Transmission
Repeatability Turbidity Scattering/
transmission
Formazin 0 NTU to 2000
NTU
Within ±2% of the reading or ±0.5
NTU, whichever is larger
2001 NTU to
4000 NTU
Within ±3% of the reading
Kaolin 0 to 1000 Within ±2% of the reading or ±0.5, whichever is larger
1001 to 2000 Within ±5% of the reading
SS Scattering/
transmission
Kaolin 0 mg/L to 1000
mg/L
Within ±2% of the reading or ±0.5mg/ L, whichever is larger
1001 mg/L to
2000 mg/L
Within ±5% of the reading
Transmission 0 mg/L to 2000
mg/L
Within ±5% of the reading or ±0.5mg/ L, whichever is larger
Linearity Turbidity Scattering/
transmission
Formazin 0 NTU to 2000
NTU
The deviation at the mid-range point of the span calibration value is within ±3% of the span calibration value or ±3 NTU, whichever is larger
2001 NTU to
4000 NTU
The deviation at the mid-range point of the span calibration value is within ±5% of the span calibration value
Kaolin 0 to 1000 The deviation at the mid-range point of the span calibration value is within ±3% of the span calibration value or ±3, whichever is larger
1001 to 2000 The deviation at the mid-range point of the span calibration value is within ±5% of the span calibration value
SS Scattering/
transmission
Kaolin 0 mg/L to 1000
mg/L
The deviation at the mid-range point of the span calibration value is within ±3% of the span calibration value or ±3 mg/L, whichever is larger
1001 mg/L to
2000 mg/L
The deviation at the mid-range point of the span calibration value is within ±5% of the span calibration value
Transmission 0 mg/L to 1000
mg/L
The deviation at the mid-range point of the span calibration value is within ±3% of the span calibration value or ±3 mg/L, whichever is larger
1001 mg/L to
2000 mg/L
The deviation at the mid-range point of the span calibration value is within ±10% of the span calibration value
Resolution Turbidity 0.01 (Range: 0 to 10) Auto decimal point or fixed decimal point can be specified
0.1 (Range: 10 to 100)
1 (Range: 100 to 4000)
SS 1 Fixed range
Absorbance 0.001 0.000 to 3.000
Concentration can be obtained by the cubic function of the absorbance.
The converted value should be within 3.000.
Transmission output Number of current outputs 2
Output type 4 mA to 20 mA DC, input/output isolation type
Load resistance 900 Ω (max.)
Linearity Within ±0.08 mA (limited to the current linearity)
Repeatability Within ±0.02 mA (limited to the current repeatability)
Error output Provided with burnout function (3.8 mA or 21 mA)
Transmission holding One of the latest value or preset value can be specified
Contact output Number of contact outputs 6
Output type No-voltage output contact
R1, R2 Contact type Relay contact, SPST (1a)
Contact capacity 250 V AC 3 A, 30 V DC 3 A (resistance load)
Contact function Selectable from among the upper limit alarm, lower limit alarm, during transmission output hold, and cleaning output
(closed at alarm operation, open under normal conditions and in a power-off state)
Setting range: Within the measurement range of turbidity or absorbance
Delay time: 0 s to 600 s
R3 Contact type Relay contact, SPST (1a)
Contact capacity 30 V DC 1 A (resistance load)
Contact function Selectable from among the upper limit alarm, lower limit alarm, during transmission output hold, and cleaning output.
(closed at alarm operation, open under normal conditions and in a power-off state)
Setting range: Within the measurement range of turbidity or absorbance
Delay time: 0 s to 600 s
FAIL Contact type Relay contact, SPDT (1c)
Contact capacity 250 V AC 3 A, 30 V DC 3 A (resistance load)
Contact function The relay coil is energized when the system is normal.
You can use the NO and NC relay contact.
Error alarms for values outside the measurement range, self-checks, and calibration errors can be set
Delay time: 0 s to 600 s
RANG1,
RANG2
Contact type Relay contact, SPST (1a)
Contact capacity 30 V DC 1 A (resistance load)
Contact function Status output of the transmission output range
Contact input Number of contact inputs 4
Contact type Open collector no-voltage “a” contact
Condition ON resistance: 100 Ω max.
Open-circuit voltage: 24 V DC
Short-circuit current: 12 mA DC max.
Contact
function
EXT1 Hold command The minus terminals of the four inputs are connected inside
EXT2 Cleaning command
EXT3, EXT4 Selection command from four
ranges using 2-bit contacts
Communication
function
Type RS-485
Signal form Two wire type, the power source of the communication is isolated from the power of measurement circuit. (The power source of transmission and communication are not isolated.)
Calibration Calibration method Zero calibration: By filtered clean water
Span calibration: Turbidity adjustment by span coefficient input
Compatible standard
substances
Kaolin, Formazin
Correction Correction method Shift correction: Shift correction to the measurement value
Gain correction: Gain correction by coefficient to the measurement value
Cleaning operation
output
(For additional
assistant cleaning)
Number of output One (For the water/air jet cleaning assistant)
Output type AC voltage output (Output voltage is the same as the power for the converter)
Contact type Relay contact, SPST (1a)
Contact capacity 250 V AC 0.5 A
Setting Cleaning cycle 0.2 h to 168.0 h
Cleaning time 30 s to 600 s
Hold time 30 s to 600 s
Timer accuracy Within 2 min per month
Cleaning start condition One of the following four methods to start cleaning is selectable
By the internal timer
By the internal timer or external cleaning command input
By the internal timer while the external permission signal is input
By the input of cleaning trigger (Cleaning input over 2 seconds triggers off the cleaning)
Self-check Sensor check error Sensor error
Converter error CPU abnormality, memory abnormality
Operating
temperature range
−20℃ to 55℃
Operating humidity
range
Relative humidity: 5% to 90% (no condensation)
Storage temperature −25℃ to 65℃
Power supply Rated power supply
voltage
100 V to 240 V AC ±10% 50/60 Hz
Power consumption 36 VA max with 100 V AC power supply
Power switch A switch is installed inside the converter.
Compatible
standards
CE marking EMC    :EN61326-1*
     Class A, Industrial electromagnetic environment
Safety   :EN61010-1
RoHS   :EN50581
      9. Industrial monitoring and control instruments
FCC rules Part 15 Class A
Structure Installation Outdoor installation type
Installation method 50 A pole-mounted or wall-mounted
Protection class IP65
Material of case Aluminum alloy (coated with epoxy-modified melamine resin)
Material of mounting
brackets
SUS304
Material of hood SUS304 (coated with epoxy-modified melamine resin)
Material of display window Polycarbonate
Display element Reflective monochrome LCD
External dimensions 180 (W)  × 155 (H)  × 115 (D) mm (excluding brackets)
Mass Mainframe: approx. 3.5 kg
Hood, bracket: approx. 1 kg

*:  The surge test specified in the EMC Directive for CE Marking shall not apply to the case when a sensor cable, transmission cable, or contact input cable is extended for 30 m or longer.

Note

Although arrestors (400 V discharge starting voltage) are provided for the contact input and communication lines, install some optimum surge absorber elements on the connection lines, depending on the surrounding environment, installation conditions, and the external devices.

Detector

Product name Industrial-use turbidity/SS detector
Model SS-150
Measurement principle Right angle scattering/transmission method or transmission method
Light source Infrared LED 860 nm
Detector Silicon photo diode
Window of light path PFA
Cell length 30 mm I.D.
Data transfer RS-485 (communication with converter)
Cleaning function Rubber blades wipe optical wall of the sensor head. (Equipped in the detector)
Actuated by stepping motor
Wiper will rotate while the cleaning command from the converter is input.
Wiper will stop at the measuring position after cleaning command is cancelled.
Sample water temperature 0℃ to 45℃ (no freezing)
Storage temperature −25℃ to 65℃
Sample water pressure 0 MPa to 0.1 MPa
Material of liquid contact PFA, POM, SUS316, FKM, EPDM, PVC, PPO, SUS304
Cable length Provided standard cable: 10 m, maximum extension distance: 50 m total 60 m.
Power supply 24 V DC 6 W supplied from HU-200TB-IM converter
External dimensions φ60 × L250 mm (excluding cables)
Mass Approx. 1.0 kg (excluding cables)
Conformation JIS K0101 ISO7027 calibration with Formazin