HE-300C

Field-installation type electrical conductivity meter (conductivity meter)

For measurements in pure water, boiler water, and pharmaceutical manufacturing processes

Combined with electrical conductivity sensors (ESH/FS series) to measure electrical conductivity and temperature of water samples. Suitable for measurement of pure water and boiler water. A lineup of sensors for pharmaceutical manufacturing processes that comply with the pharmacopoeias (USP, EP, JP) of various countries are also offered.

COND

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Diverse lineup of sensors available

Sensors that are tailored to match the sample characteristics and applications are proposed.

Highly robust

Required accessories and wiring for installation possess the ruggedness appropriate for industrial applications, ensuring they can withstand harsh environmental conditions.

Product name Conductivity meter
Model HE-300C
Sensors in combination 2-electrode method conductivity sensor
ESH and FS Series sensor of cell constant 0.01/cm, 0.1/cm, or 1.0/cm
Measurable
range
(*Note 1)
Cell constant (/cm) 0.01
(ESH-001)
0.1
(ESH/FS-01)
1.0
(ESH-1)
  Unit
Conductivity μS/cm 0.000 to 2.000
0.00 to 10.00
0.000 to 2.000
0.00 to 20.00
0.0 to 100.0
0.0 to 200.0
0 to 1000
mS/m 0.0000 to 0.2000
0.000 to 1.000
0.0000 to 0.2000
0.000 to 2.000
0.00 to 10.00
0.00 to 20.00
0.0 to 100.0
TDS
conversion
mg/L 0.00 to 2.00
0.0 to 10.0
0.00 to 2.00
0.0 to 20.0
0 to 100
0 to 200
0 to 1000
Temperature 0 to 100 (display range: −10℃ to 160℃)
Display
resolution
Conductivity
/ TDS conversion
As shown in the measuring table.
Temperature 0.01℃
Performance Conductivity Repeatability Within ±0.5% of the full scale (response for equivalent input)
Within ±5% of the full scale in the range of 2000 μS/cm or 200 mS/m for FS-01 sensor
Linearity Within ±0.5% of the full scale (response for equivalent input)
Within ±5% of the full scale in the range of 2000 μS/cm or 200 mS/m for FS-01 sensor
TDS conversion Repeatability Within ±1.5% of full-scale value (response for equivalent input)
Linearity Within ±1.5% of full-scale value (response for equivalent input)
Temperature Repeatability ±0.1℃(response for equivalent input)
Linearity ±0.5℃(response for equivalent input)
Transmission
output
Output type 4 mA to 20 mA DC input/output insulation type (two-wire transmission type)
Load resistance Maximum: 600 Ω. Case of 24 V DC power supply (*Note 3)
Linearity Within ±0.08 mA (output only)
Repeatability Within ±0.02 mA (output only)
Output range Conductivity: Free setting within a measuring range.
Occasional out for error Hold or burnout to either 3.8 mA or 21 mA
Transmission hold In the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact
input
Number of input points 1
Contact type No-voltage a-contact
Conditions ON resistance: 40 Ω max.
Open voltage: 1.2 VDC
Short-circuit current: 21 mA DC max.
Contact function External input for transmission holding.
Temperature
compensation
Temperature element Platinum resistor: 1 kΩ (0℃)
Temperature compensation
  • Temperature characteristic of NaCl (reference temperature: 5℃ to 95℃)
  • Arbitrary temperature coefficient input (reference temperature: 5℃ to 95℃; temperature coefficient: ±3%/℃)
  • No temperature compensation
    One of the three ways of compensation is selectable.
    (In the de-ionized water area, the temperature compensation for pure water is automatically performed in both NaCl and arbitrary temperature coefficient compensation settings.)
Temperature compensation
range
0℃ to 100℃ (However, the compensation calculation is extended lower than 0℃ or higher than 100℃.)
Temperature calibration 1 point calibration comparing to reference thermometer
Calibration Conductivity Based on parameter input of coefficient for the sensor cell constant.
TDS conversion Conversion by user-defined coefficient value
(TDS calculation is 0.30 to 1.00 times of sample conductivity by μS/cm.)
Self-check Sensor diagnosis error Temperature sensor short-circuit, temperature sensor disconnection,
and out of the temperature measurement range
Meter error CPU error, ADC error, and memory error
Operating temperature range −20℃ to 60℃ (without freeze)
Operating humidity range Relative humidity: 5% to 90% (without condensation)
Storage temperature −25℃ to 65℃
Power
supply
Rated voltage 24 V DC (operating voltage range: 21 V to 32 V DC) (*Note 3)
Power consumption 0.6 W max.
Compatible
standards
CE marking EMC directive: EN61326-1
RoHS Directive: EN50581
  Immunity
Industrial
electromagnetic
environment
Electostatic discharge IEC61000-4-2
Electromagnetic field of radiated radio frequency IEC61000-4-3
Electric fast transient/burst IEC61000-4-4
Surge IEC61000-4-5
(*Note 2)
Conducted interference induced by radio frequency IEC61000-4-6
Emission
Class A
Radiated disturbance CICPR 11 CLASS A
Noise terminal voltage CISPR 11 CLASS A
FCC Rules Part15 CLASS A
Structure Installation Outdoor installation type
Installation method 50 Mount on 50 A pole or wall
Protection code IP65
Case material Aluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittings SUS304
Material of hood SUS 304 stainless steel (coated with epoxy-denatured melamine resin)
Material of window Polycarbonate
Display element Reflective monochrome LCD
External dimensions 180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets)
Mass Main body: Approx. 2.8 kg; cover and mounting brackets: Approx. 1 kg

Note

  • Note 1: Only the sanitary sensor (FS-01 series) is applied.
  • Note 2: Conductivity measurement and TDS conversion measurement cannot be selected at the same time.
  • Note 3: The maximum load resistance that can be connected is decided depending on the powersupply voltage.

 

Relation between power supply voltage and load resistance

Relation between power supply voltage and load resistance

  • Note 4: When the sensor cable, the transmission cable, or the contact input cable is extended by 30 m or more, the surge test under the EMC Directive for CE marking is not applied.
  • Note 5: An arrester (spark over voltage: 400 V) is implemented for transmission output, contact input, and communication. However, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.