TPNA-500

Automatic total nitrogen/phosphorus monitor

The latest model combines 60 years of water measurement technology to achieve reduced life cycle costs, compliant regulations.

The future of total nitrogen and total phosphorus measurement not only involves compliance with regulations but also has the potential to contribute to proactive enhancements of corporate values. The future of measurement, regulatory compliance, and enhancement of corporate value will be pursued by proposing technological innovation, improvement, and LCC (Life Cycle Cost) reduction.

TP/TN

Division: Water Pollution
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Extension of reagent lifespan (twice as long as conventional)

The stability of potassium persulfate has been improved through the development of dedicated reagents. The reagent replacement cycle has been extended from one month to two months. *1
*1: Depends on reagent storage and installation conditions (ambient temperature) (compared to our previous products)

Power consumption has been reduced by approximately 30%, and the amount of pure water and waste liquid has been halved

Reduction of costs (including labor) and environmental impact associated with electricity consumption, use of pure water, and waste liquid treatment (compared to previous estimates).

Adoption of a robust weighing system resistant to contamination

Adoption of a new weighing method (meniscus weighing method) and enhanced automatic cleaning function (optional), while maintaining the reagent volume reduction (1/10) achieved with our previous models. Contributes to extending the maintenance cycle.

Reevaluation of existing equipment

Improved have been made to specifications to make the equipment more reliable by reducing the use of mechanically driven components such as solenoid valves.

General
specifications
Equipment name Automatic total nitrogen/phosphorus monitoring system
Model TPNA-500
Dimensions *1 460 (W) mm × 385 (D) mm × 1600 (H) mm
Mass Approx. 87 kg
Power supply 100 V to 240 V AC ±10%, 50 Hz/60 Hz
Power consumption 100 V to 240 V AC: Approx. 250 VA
Performance Measurement
range *2
Total nitrogen: 0 mg N/L to 2/5/10/20/50/100/200/500/1000 mg N/L
Total phosphorus: 0 mg P/L to 0.5/1/2/5/10/20/50/100/250 mg P/L
Repeatability Within ±3% of the full scale
(Total nitrogen full scale: 50 mg N/L or less, total phosphorus full scale: 10 mg P/L or less)
Within ±5% of the full scale
(Total nitrogen full scale: 100 mg N/L to 1000 mg N/L, total phosphorus full scale: 20 mg P/L to 250 mg P/L)
Both with standard solution
Measurement
principle
Total nitrogen: Alkaline potassium persulfate, ultraviolet oxidative dissolution -
ultraviolet absorption spectrophotometry
Total phosphorus: Potassium persulfate, ultraviolet oxidative dissolution -
Molybdenum blue spectrophotometric method
Number of
measurement
points
One point (standard), two points (option)
Number of
measurement
ranges *3
One range (standard), 2 ranges (option)
Measurement
cycle
60 min
Installation
conditions
Ambient
temperature
2℃ to 40℃
Ambient humidity Relative humidity 85% or less (no condensation)
Ambient
environment
  • Indoor installation type
  • Transient overvoltage of main power supply: Overvoltage Category II (EN61010- 1), Pollution Degree 2
  • Flat surface that is not subject to vibration and external shock
  • No dust, mist, or corrosive gas is present in the atmosphere
  • Under atmospheric pressure
  • No direct sunlight
  • Well-ventilated area
Measurement
conditions
Sample water
conditions
*4, *5, *6,*7
Temperature 2℃ to 40℃ (no freezing)
Flow rate
  • 0.5 L/min to 5 L/min (when OF-3 is used)
  • 5 L/min to 20 L/min(when OF-30 is used)
Pure water/tap
water supply
conditions*8
Amount of usage 43 L/month (depending on the measurement range)
Water quality
  • When the customer supplies pure water, use the water of class
    A2 or above, as specified in JIS K0557: 1998, which does not contain nitrogen and phosphorus compounds.
  • When a pure water unit is used, the tap water criteria defined by the Water Supply Act must be met.
Supply method Supply by the tank (standard) or continuous supply by the pure water equipment (option)
Supply pressure 0.1 MPa to 0.5 MPa
Input/output
specifications
Analog output Number of
points
Standard: 4 points
Option (3-component specification/2-point meter specification): 8 points
Type Standard: 4 mA to 20 mA DC, 0 mA to 16 mA DC
(Set to 4 mA to 20 mA DC at the factory. It can be switched to 0 mA to 16 mA DC from the screen.)
Option: 0 V to 1 V DC, 1 V to 5 V DC
Description Standard: TN concentration, TP concentration, TN load, TP load, flow rate per hour
Option (3-component specification): COD concentration, COD load, COD instantaneous concentration
Note: The COD output settings depend on the equipment
conditions.
Output
impedance
500Ω or less
Contact output Number of
points
Standard: 16 points
Option (3-component specification/2-point meter specification): 22 points
Type No-voltage output contact
Form Isolated output
Output capacity 250 V AC 3 A
30 V DC 3 A (Only 30 V DC 3 A is available for the “Maintenance” contact)
Status output Meas., Calibration, Standby, Maintenance, SYNCHRONOUS IDLE
1, SYNCHRONOUS IDLE 2, Power, etc.
Alarm output Hi limit alarm (TP, TN, COD), H.Hi limit alarm (TP, TN, COD), Load limit alarm (TP, TN, COD), Sample Lack, Lack Blank, Lack Reag.,
Lack Span, Full Drain Tank, Total Alarm 1 to 6, various meter errors, etc.
Analog input Number of
points
Standard: 1 point
Option (3-component specification/2-point meter specification):
2 points
Type Standard: 4 mA to 20 mA DC
Option: 1 V to 5 V DC
Description Flow rate (full scale can be set as desired), UV/COD meter
concentration
Contact input Number of
points
Standard: 9 points
Option (3-component specification/2-point meter specification): 18 points
Type No-voltage a contact input (open collector is available)
Form Isolated input: Common to the (-) side
ON resistance 100 Ω (max.)
Open-circuit
voltage
26 V DC (max.)
Short-circuit
current
13 mA DC (max.)
Function Meas. Start, Cal. Start, Modify Date, Line Select,
Flow Mainte, Flow Err., Flow Power, Flow No Drain,
UV/COD Mainte, UV/COD Err., UV/COD Power, etc.
Equipment
function
Display Touch panel type liquid crystal (color)
Load calculation function Standard: TP/TN load calculation function is provided
Option: COD and 2-point meter load calculation function are provided
Option Turbidity compensation function is provided (TN)
(Turbidity compensation wavelength: 275 nm)
Memory function Measurement value: Memory for over 1 year (including the daily and hourly reports)
Alarm history: 500 events
Calibration history: 100 events
Operation history: 100 events
Note: USB external memory device can be used.
Communication function Standard: RS-485
Option: RS-232C
Communication protocol Modbus
Printer Provided (recording paper width: 58 mm) With automatic take-up equipment
Reagent Reagent A 0.8 mL per measurement (capacity: 1000 mL)
Solution B 0.2 mL per measurement (capacity: 250 mL)
Solution C 0.1 mL per measurement (capacity: 250 mL)
Solution D 0.1 mL per measurement (capacity: 250 mL)
Solution E 0.19 mL per measurement (capacity: 250 mL)
Solution F 0.2 mL per measurement (capacity: 250 mL)
Span solution 50 mL per measurement (capacity: 1000 mL)
Replacement cycle One month (only under certain storage/usage environment)

*1:  Compatible with older products: Model TPNA-300, PN-100 series channel base (standard)
*2:  Full-scale values for both total nitrogen and total phosphorous must be greater than or equal to 0.2 and less than or equal to 100
*3:  When the first range is set to low concentration, the following condition must be met for the same measurement component:
    1 ≦ Second range full scale/first range full scale ≦ 50
*4:  The chemical cleaning function for the sample line can be selected according to the nature of the sample
*5:  The provided overflow tank can be installed in either the left/right side or rear of the main unit
*6:  The distance from the provided overflow tank to the main unit should be within 1 m
*7:  The sample must not contain the components that could affect the measurement (metallic ions, etc.)
*8:  When tap water is supplied, flush the pipes by running the water for approximately 30 minutes before supplying.

Note

Perform the installation work while taking the features of the analyzer into consideration. Also, ensure an environment in which the equipment can operate normally. If the equipment is not operating normally due to impairment, operating conditions, or other unexpected events, take appropriate action so that no damage to the equipment will occur.