| XC Mode | Standard Mode | ||
|---|---|---|---|
| Measurement Principle | High-frequency induction furnace combustion – infrared absorption method | ||
| Sample Mass | Max 0.50 or less than 0.49 ± 0.01 g | Less than 1.00 ± 0.10 g | |
| Measurement Time (Average) | Approx. 160 seconds | Approx. 60 seconds | |
| Carbon | Measurement Range *2 | 1.0% - 20.0% | 0.6 ppm - 10.0% |
| Reproducibility *3 | RSD ≤ 0.15% | σ ≤ 0.3 ppm (Sample less than 20 ppm), σ ≤ 1 ppm or RSD ≤ 0.5% (Sample more than 20 ppm) | |
| Sulfur | Measurement Range *2 | - | 0.6 ppm - 1.0% |
| Reproducibility *3 | - | σ ≤ 0.3 ppm (Sample less than 20 ppm), σ ≤ 1 ppm or RSD ≤ 0.75% (Sample more than 20 ppm) | |
Utility
| Carrier Gas | Oxygen (Purity: 99.5% or higher, 0.30 - 0.33 MPa) |
|---|---|
| Operation Gas | Nitrogen (Purity: 99.5% or higher, 0.40 - 0.45 MPa) |
Dimensions (W x D x H mm) and Mass (kg)
| Main Unit | 500 x 725 x 710 mm/approx. 134 kg*4 |
|---|---|
| Purifier/Oxidizer | 160 x 560 x 512 mm/approx. 28 kg |
| XC Unit | 260 x 560 x 513 mm/approx. 37 kg |
*1 Varies depending on the sample, measurement conditions, and options.
*2 Maximum range is extended by reducing the sample mass.
*3 Guaranteed values apply and are limited to the following conditions: When using reliable standards and using official consumables (accelerators, crucible, reagents, others).
*4 Excludes protrusions on the rear panel.
Available Options
- Auto Sampler
Up to 20 samples can be loaded. The system automatically places crucibles in the furnace and disposes of them after measurements. Measurements can also be manually interrupted when necessary.

- Halogen Trap Unit
HT-03 Traps halogen derived from sample, allowing analysis with minimized influence.can be loaded. The system automatically places crucibles in the furnace and disposes of them after measurements.