HC-200N

Ammonia/Nitrate Nitrogen Meter

Ammonia nitrogen and nitrate nitrogen concentrations in the tank are measured in real time with an ion electrode method sensor.

Reduces power consumption and costs at sewage treatment plants and other water treatment facilities. The proprietary technology enables high durability and long-term stability.

 

NH4-NNO3-N

事业部: 过程与环境
产品分类: 过程水体
制造商: HORIBA Advanced Techno Co., Ltd.

Simultaneous measurement of two components

Enables simultaneous measurement of ammonia nitrogen and nitrate nitrogen concentrations, leading to reduced power consumption and costs at water treatment facilities such as nitrification-denitrification tanks.

Long-term stability

The sensor's unique protective structure and optimized internal solution enable stable measurements. 

Easy maintenance

Replaceable chip without the need for tools. No need to call manufacturer maintenance.

Prevention of contamination for increased longevity

HORIBA's proprietary protective film provides sensor durability. Combined with an ultrasonic cleaner, antifoulding properties are improved.

Reduced risk of sudden operational shutdowns

The self-diagnostic function of the sensor enables the immediate detection of anomalies.
*Ammonia nitrogen only.

 

HC-200N Specifications

Product name Ammonia/Nitrate Nitrogen Meter
Converter model HC-200N

 

Ammonia nitrogen

Sensor model AM-2000-A: Ammonia nitrogen sensor
Sensor tip model 7691: Ammonium ion tip, 7692: Potassium ion tip for compensation, 7211: Reference tip
Measurement range Concentration: 0 mg/L to 1000 mg/L (display range: 0 mg/L to 2000 mg/L)
Temperature: 0℃ to 40℃ (display range: -10℃ to 110℃)
Display resolution Concentration: 0.01 mg/L: 0.00 mg/L to 10.00 mg/, 0.1 mg/L: 0.0 mg/L to 100.0 mg/L, 1 mg/L: 0 mg/L to 1000mg/L
Temperature: 0.1℃
Repeatability Concentration: ±3% or ±0.2 mg/L, whichever is greater for the readings (standard solution at 25℃)
Temperature: ±0.3℃(equivalent input)
Temperature 
compensation
Compatible temperature element: platinum resistor: 1 kΩ (0℃)
Temperature measurement range: 0℃ to 40℃
Temperature calibration function: 1 point calibration for comparison with the reference thermometer
Potassium ion
concentration
compensation
Compensation range: the potassium ion concentration is not more than 10 times the ammonium ion concentration and 1000 ppm or less.
Compensation error: ±20%(reading)
Measurable pH range pH 4.0 to 8.5
Calibration Reference solution calibration of the ammonia nitrogen (2points)
Type of the reference solution: select 1 mg/L to 10 mg/L or 10 mg/L to 100 mg/L.
Additional functions: automatic defect calibration judgment, calibration history
Adjustment Sample adjustment using hand analysis (1 point, 2 points*1), standard curve input function (primary system)
Additional functions: adjustment history (number of days elapsed from the previous adjustment, sample adjustment coefficient)
Abnormal detection Calibration error, sensor diagnosis error, converter error

*1 When AM-2000 is connected, the 2-point sample adjustment function becomes unavailable.

 

Nitrate nitrogen

Sensor model NO-2000: Nitrate nitrogen sensor
Sensor tip model 7693: Nitrate ion tip, 7694: Chloride ion tip for compensation, 7211: Reference tip
Measurement range Concentration: 0 mg/L to 1000 mg/L (display range: 0 mg/L to 2000 mg/L)
Temperature: 0℃ to 40℃ (display range: -10℃ to 110℃)
Display resolution Concentration: 0.01 mg/L: 0.00 mg/L to 10.00 mg/, 0.1 mg/L: 0.0 mg/L to 100.0 mg/L, 1 mg/L: 0 mg/L to 1000mg/L
Temperature: 0.1℃
Repeatability Concentration: ±3% or ±0.2 mg/L, whichever is greater for the readings (standard solution at 25℃)
Temperature: ±0.3℃(equivalent input)
Temperature 
compensation
Compatible temperature element: platinum resistor: 1 kΩ (0℃)
Temperature measurement range: 0℃ to 40℃
Temperature calibration function: 1 point calibration for comparison with the reference thermometer
Chloride ion
concentration
compensation
Compensation range: when the chloride ion concentration is 100 times greater than the nitrate ion concentration, a correction factor must be applied to account for interference.
Compensation error: ±20%(reading)
Measurable pH range pH 2.0 to 12.0
When the pH is between pH 2.0 to pH 4.0 or pH 9.0 to pH 12.0, use sensor cap AM-CAP-PVC.
Chloride ion tip can be used in the pH 2.0 to pH 8.0. However, when used at pH 2.0 to pH 4.0, the lifetime of the tip may be shortened.
Measurement inhibitors Component: if the surfactant concentration is 10mg/L or higher, sensor performance may degrade, and measurement may not be possible. 
Calibration Reference solution calibration of the ammonia nitrogen (2points)
Type of the reference solution: select 1 mg/L to 10 mg/L or 10 mg/L to 100 mg/L.
Additional functions: automatic defect calibration judgment, calibration history
Adjustment Sample adjustment using hand analysis (1 point, 2 points), standard curve input function (primary system)
Additional functions: adjustment history (number of days elapsed from the previous adjustment, sample adjustment coefficient)
Abnormal detection Calibration error, sensor diagnosis error, converter error

 

Common specifications (Converter)

Transmission output 3 points, 4 mA to 20 mA DC, input/output isolation type, 900Ω(max)
Select 3 items from below.
Ammonia nitrogen concentration: arbitrary setting within the measurement range is possible.
Nitrate nitrogen concentration: arbitrary setting within the measurement range is possible.
Chloride ion concentration: arbitrary setting within the measurement range is possible.
Temperature reading of the ammonia nitrogen meter: arbitrary setting within the range from 
⁻10℃ to 110℃ is possible.
Temperature reading of the nitrate nitrogen meter: arbitrary setting within the range from 
⁻10℃ to 110℃ is possible.
Contact output 5 points, No-voltage contact output, relay contact, SPST(1a),SPDT(1c)
 R1,R2: relay contact SPST(1a)
           Contact function: selectable from among the upper limit alarm, lower limit alarm, during
           transmission output hold, and cleaning output (closed at alarm operation, open under 
           normal conditions and in a power off state).
 FAIL: relay contact SPDT(1c)
           Contact function: error alarms for values outside the measurement range and self 
           check output can be set. 
           Delay time: 0 seconds to 600 seconds
 R3,R4: relay contact SPST(1a)
Cleaning output 2 points, Voltage contact output (connection power supply voltage output)
Relay contact, SPDT(1c), SPST(1a)
Contact input 1 point, Contact type: open collector no-voltage “a” contact
Communication function Type: RS-485
Signal form: Two wire type, the power source of the communication is isolated from the power of measurement circuit.(The power source of transmission and communication are not isolated.)
Operating temperature range ⁻20℃ to 55℃(no freezing)
Operating humidity range Relative humidity: 5% to 90%(no condensation)
Storage temperature ⁻25℃ to 65℃ 
Power supply 100V to 240V AC ±10% 50/60Hz, Power consumption: 28 VA(max)
Compatible standards CE marking, FCC Rules
Structure Outdoor installation type, IP65, 50 A pole-mounted or wall-mounted, material of case: aluminium alloy, material of the mounting brackets and hood :SUS304
External dimensions 180(W)×155(H)×115(D)mm (excluding mounting brackets)
Mass Mainframe: approx. 3.5kg, hood, mounting bracket: approx. 1kg

 

Ammonia nitrogen sensor

Model AM-2000-A
Measurement range Temperature: 0℃ to 40℃ (no freezing)
Measurement depth 0m to 10m
Material SUS316*², FKM, PVC
Mass Approx. 2.7kg (including 10m cable)

*2 The instrument may corrode when used in seawater or in samples with a high chloride concentration.

 

Nitrate nitrogen sensor

Model NO-2000
Measurement range Temperature: 0℃ to 40℃ (no freezing)
Measurement depth 0m to 10m
Material SUS316*³, FKM, PVC, POM
Mass Approx. 2.7kg (including 10m cable)

*3 The instrument may corrode when used in seawater, in samples with a high chloride concentration, or in samples with a pH below 4.