NOx | SO2 | CO | CO2 | O2 | CH4 | ||
---|---|---|---|---|---|---|---|
Standard | PG-320 |
| ● | ● | |||
PG-325 | ● | ● | |||||
PG-330 | ● | ● | ● | ||||
PG-335 | ● | ● | ● | ||||
PG-337 | ● | ● | ● | ||||
PG-340 | ● | ● | ● | ● | |||
PG-350 | ● | ● | ● | ● | ● | ||
CH4 | PG-324 | ● | ● | ||||
PG-344 | ● | ● | ● | ● |
Analysis Principle | |
---|---|
NOx | Cross-Flow Modulation Chemiluminescence Detection Method |
SO2 ,CO, CH4 | Cross-Flow Modulation Non-Dispersive Infrared Absorption Method |
CO2 | Non-Dispersive Infrared Absorption Method |
O2 | Galvanic, Zirconia or Paramagnetic Method |
Ranges (Standard models) | |
---|---|
NOx | 0-25/50/100/250/500/1000/2500 ppm |
SO2 | 0-200/500/1000/3000 ppm |
CO | 0-200/500/1000/2000/5000 ppm |
CO2 | 0-10/20/30 vol% |
O2 | 0-5/10/25 vol% * |
Ranges (CH4 models) | ||
---|---|---|
CH4 | 0-2000/5000ppm | |
CO | 0-2000/5000ppm | |
CO2 | 0-5/10/20vol% | |
O2 | 0-5/10/25vol%* |
Standard Model | CH4 Model | |
---|---|---|
Repeatability | ±0.5% of full scale (NOx:≧100ppm range / CO: ≧1000ppm range) ±1.0% of full scale (Except above) | ±1.0% of full scale |
Linearity | ±2.0% of full scale | |
Drift | ±1.0% of full scale / day (For SO2 analyzer only: ±2.0% of full scale / day) | |
Response Time (Td+T90) | Analyzers except SO2 analyzer: 45 sec. or less (From sample inlet) SO2 analyzer: 180 sec. or less (From sample inlet) | |
Sample Gas Flow Rate | Approx. 0.5L/min. | |
Display | Measurement (3 or 4 digit display), range flow rate, etc. | |
Output | DC 4-20 mA (non-insulated) or 0-1 V (non-insulated)[optional] / Ethernet | |
Warm-up Time | 30 min. ±2.0% of full scale / 2 hours | |
Data Saving | SDTM memory card / SDHCTM memory card | |
Ambient Temperature | 0°C-40°C <32°F-104°F> | |
Ambient Humidity | Maximum 80% R.H. | |
Power | AC 100V to 240V ±10% (maximum 250 V), 50/60 Hz | |
Power Consumption | Approx. 160VA in a steady state, maximum 220VA | |
Dimentions | (With side guards) 300(W) x 520(D) x 260(H) mm <11.8"(W) x 20.5"(D) x 10.2"(H)> (Without side guards) 260(W) x 520(D) x 260(H) mm <10.2"(W) x 20.5"(D) x 10.2"(H) | |
Weight | Approx. 13kg - 15kg <29lb ~ 33lb> | |
Environmental rating | IP42(Optionally available) | |
Sample Gas Condition | Temperature: Ambient temperature |
・SD is a trade mark for SD-3C, LLC.
・TAKE GREAT CARE WHEN HANDLING SAMPLE GASES CONTAINING TOXIC OR FLAMMABLE GASES. TAKE MEASURES SUCH AS PROVIDING ADEQUATE VENTILATION, INSTALLING GAS DETECTORS, AND REMOVING IGNITION SOURCES IN THE WORKING AREA.
・THE PG-300 SERIES IS NOT EXPLOSION-PROOF. DO NOT USE THIS PRODUCT IN A HAZARDOUS LOCATION OR FOR MEASUREMENT OF SAMPLE GASES IN EXPLOSIVE ATMOSPHERES (MIXTURE OF A COMBUSTIBLE GAS AND AIR WITHIN THE FLAMMABILITY LIMITS). HORIBA, LTD. AND ITS AFFILIATES ARE NOT LIABLE FOR EMERGENCIES CAUSED BY LEAKAGE OR MISHANDLING OF SUCH GASES.