SLIA-5000

SILICA ANALYZER

Quick and automatic measurement for boiler water, and trace amount of silica present in pure water.

Management of silica (SiO2) is paramount in preventing the formation of scales. This is because scales present in turbines will reduce the efficiency of power generation, especially for thermal power plants and nuclear power plants. By combining HORIBA's many years of track records and technological know-how, the silica analyzer has been reborn with conventional stability, reliability and maintainability.

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.
  • ±2% F.S. repeatability with measurement time of 5 minutes
    Capable of obtaining high-speed, and highly repeatable measurement results.
  • Wide measurement range
    Measurement is possible from high sensitivity range of 0 to 10μg/L (ppb) to high concentration range of 0 to 5.0mg/L (ppm).
  • Self-diagnosis function
    Self-diagnosis of operation abnormalities such as water sample or reagent supply error, calibration error, light source error, and in-cell temperature error.
  • Automatic calibration function
    Zero calibration is performed automatically for each measurement. Span calibration can also be preset to perform at specific timings.
  • Automatic switching of measuring cycle
    Switching of measuring cycle automatically saves energy and reagents because measuring cycle is automatically shortened when the sample concentration exceeds the set concentration level, and automatically lengthened when the sample concentration goes below the set level.
  • Improved operability
    Equipped with color LCD touch panel.
General
specifications
Equipment nameSILICA ANALYZER
ModelSLIA-5000
Dimensions600 (W) mm × 750 (H) mm × 1600 (D) mm
MassApproximately 130 kg
Power supply100 V AC ±10%, 50/60 Hz
(Option) 110 V AC/115 V AC/120 V AC/220 V AC/230 V AC/240 V AC/
240 V AC±10%, 50/60 Hz
Power
consumption
Approximately 500 VA
PerformanceMeasurement
range
  • 0 μg/L to 10 μg/L/0 μg/L to 100 μg/L
  • 0 μg/L to 50 μg/L/0 μg/L to 500 μg/L
  • 0 μg/L to 100 μg/L/0 μg/L to 1000 μg/L
  • 0 μg/L to 200 μg/L/0 μg/L to 2000 μg/L
  • 0 mg/L to 0.50 mg/L/0 mg/L to 5.00 mg/L
    (Either is specified when receiving the order)
Resolution
(Unit setting: μg/L)
0.1 μg/LData<100 μg/L
1 μg/L100 μg/L≦Data≦5000 μg/L
Resolution
(Unit setting: mg/L)
0.0001 mg/LData<0.1 mg/L
0.001 mg/L0.1 mg/L≦Data≦5.0 mg/L
Repeatability±2% of the full scale
Measurement
principle
Molybdenum blue method
Number of
measurement
points
1 point to 4 points (specified when receiving the order)
Measurement time5 minutes
Measurement cycleSpecified as desired between 5 minutes and 999 minutes
CalibrationZeroEach measurement
SpanEqualizing filter method
Reagent
replacement cycle
Once a week (reagent storage: 30℃ or below)
Installation
conditions
Ambient
temperature
0℃ to 45℃
Ambient humidityRelative humidity 85% or less (no condensation)
Ambient
environment
  • Indoor installation type
  • Transient overvoltage of main power supply: Overvoltage Category II
    (EN61010-1), Pollution Degree 2
  • Flat surface that is not subject to vibration and external shock
  • No dust, mist, or corrosive gas is present in the atmosphere
  • Under atmospheric pressure
  • No direct sunlight
  • Well-ventilated area
Measurement
conditions
Sample conditionsTemperature2℃ to 40℃ (36°F to 104°F)
Flow rate0.5 L/min to 1.0 L/min
Pressure0.02 MPa to 0.20 MPa
pH value11.0 or less
Coexisting phosphate
ion concentration
20 mg/L or less
However, no phosphate ions are permissible for the 0 μg/L to 10 μg/L/
0 μg/L to 100 μg/L range (ultrapure water specification)
External
input/output
Analog outputConcentration output
4 mA DC to 20 mA DC or 0 mA DC to 16 mA DC, selection setting in the operation section
Marker/hold output, selection setting in the operation section
Maximum load of 500 Ω or less, input/output isolation type, nonisolated between channels
Option: 0 V DC to 1 V DC, 1 V DC to 5 V DC
Contact outputStandarda-contact output (b-contact can be specified when receiving the order)
Contact rating250 V 3 A AC, 30 V 3 A DC
Measurement line
shared
Meas., Standby, Maintenance, External control,
Sync. Signal, Equipment Err., Power, etc.
Measurement line
individual
Line, Sample Lack, Abnormal concentration,
etc.
Contact inputNo-voltage a-contact input (open collector is available)
Isolated input: Common to the (-) side
ON resistance100 Ω (max.)
Open-circuit voltage26 V DC (max.)
Short-circuit current13 mA DC (max.)
Measurement line
shared
Meas. Start, Cal. Start
Measurement line
individual
MEAS. LINE SET, MEAS. RANGE SET
Equipment
function
DisplayColor touch panel type liquid crystal
Memory functionMeasurement valueMemory for over 1 year
Alarm history500 events
Calibration history100 events
Operation history100 events
USB external memory device can be used
OptionFilter for sample line
(For Line 1 to Line 4)

By installing about 5 μm cartridge filter in the upstream of the overflow tank, it is possible to prevent iron rust, etc. from entering the equipment.

  • Up to 2 filters can be built into the product.
  • If the optional filter is selected for the product of 3 or 4 measurement
    points, the filter is not built into the product but is delivered individually as a single filter.
  • If a step-down transformer is installed (power supply voltage is other
    than 100 V AC), a filter cannot be built into the product.
Flowmeter for sample
line
(For Line 1 to Line 4)
By adding this, you can accurately measure the sample supply flow rate.
  • When installing the analyzer, give careful consideration to its special characteristics and be sure to provide an environment where the equipment can operate properly.
  • Take appropriate action so that no damage will occur even when the equipment no longer operates correctly due to deterioration or the operational conditions of the equipment, or any unexpected phenomenon.

Request for Information

Do you have any questions or requests? Use this form to contact our specialists. * These fields are mandatory.

Related products

More Aqualog - Water Treatment Plant Analyzer

Automated Organic Analysis & Early Warning Sentinel

More Aqualog Datastream

For Instant Water Quality Reports

More Aqualog Datastream

For Instant Water Quality Reports

More CS-100 Series

Stand-alone Type Chemical Concentration Monitor

More CS-100F1 Series

Fiber Optic Type Chemical Solution Concentration Monitor

More CS-600F

Fiber Optic Type Chemical Concentration Monitor

More CS-620F

Optical Fiber Type Hot Phosphoric Acid Concentration Monitor

More CS-700

High Precision, High Stability Chemical Concentration Monitor

More CS-900

Non-Contact Chemical Concentration Monitor

More EG-100

Washwater Monitoring System

More HC-200F

Field-installation Type Simplified Fluoride Ion Concentration Meter(Four-Wire Transmission)

More HC-200NH

Field-installation Type Ammonia Nitrogen Meter(Four-Wire Transmission)

More HC-300F

Field-installation Type Simplified Fluoride Ion Concentration Meter(Two-Wire Transmission)

More HD-960L

Dissolved Oxygen Monitor in Low Concentration

More HD-960LR

Dissolved oxygen concentration monitor series for semiconductor manufacturing

More HE-480C-GC

Carbon Sensor Conductivity Meter (Low concentration type)

More HE-960CA

Citric Acid Monitor

More HE-960H-TM

Wide range TMAH Concentration Monitor

More HE-960H-TM-S

High Precision TMAH Concentration Monitor

More HE-960HC

Carbon Sensor Conductivity Meter (High concentration type)

More HE-960LC

Carbon Sensor Conductivity Meter

More HE-960LF / FS-09F-1/2

Flat Carbon Sensor Conductivity meter

More HE-960R-GC(W)

Carbon Sensor Resistivity Meter

More HE-960RW-GC(W)

2-Channel Resistivity Meter

More HE-960TM

TMAH Conductivity Meter. Perfect for Controlling TMAH Developer Concentrations

More HF-960EM

In-line Sensor & Auto Range Switching Concentration Monitor

More HF-960H

HF / HCl Concentration Monitor

More HF-960M

Low Concentration Type HF/HCl/NH3 Concentration Monitor

More HF-SP-CU

2-channel High Precision HF Concentration Monitor

More HT-110

Total Organic Carbon Monitor

More HZ-960

Dissolved Ozone Monitor

More ICP Neo

ICP Software

More OPSA-150

Organic Pollutant Monitor

More SLIA-300

Highly Sensitive Silica Monitor

More TPNA-500

Automatic Total Nitrogen/Phosphorus Monitoring

More Ultima Expert LT

Affordable high performance ICP-OES

More UP-100

Micro Volume pH Monitor

More CODA-500

Automatic COD Monitor

More EG-100

Washwater Monitoring System

More SLIA-300

Highly Sensitive Silica Monitor

More TPNA-500

Automatic Total Nitrogen/Phosphorus Monitoring

More TW-100

Water Distribution Monitor

More TW-150

Online Water Quality Monitor

More U-50

Multi-parameter Water Quality Checker

More Yumizen C1200 AL

Clinical Chemisty Analyzer