Panel-mount Type pH Meter (Four-Wire Transmission)
HP-480 connects pH sensor and measures pH and temperature in the sample water. Suitable for monitoring and control of various production processes and wastewater treatment, etc. Many types of electrodes are available to match the sample properties.
pH: 0 - 14
Temperature: 0 - 100 ℃
Product name | Industrial pH meter | ||
---|---|---|---|
Model | HP-480 | ||
Measuring range | 0 pH to 14 pH: 0.01 pH resolution. Temperature 0℃ to 100℃: 1℃ resolution (selectable display). | ||
Transmission output | 4 mA to 20 mA DC: input/output isolated type Maximum load resistance: 900 Ω | ||
Transmission output range | Free | ||
Repeatability and linearity | ±0.05 pH, ±0.08 mA (for transmission output range of 0 pH to 14 pH) | ||
Contact output | Outputs: 2 points Alarm contact output (R1 and R2) Contact type: relay contact, SPDT (1c) Contact rating: 240 V AC, 3 A and 30 V DC, 3 A (resistance load) Contact function: selectable from upper/lower limit operation (ON/OFF control), alarm, and maintenance. | ||
Calibration function | Two-point automatic calibration or manual calibration. Two-point automatic calibration: with automatic validation of the electrode characteristics. Types of standard solution: pH 2, pH 4, pH 7, pH 9 and pH 10 (JIS) Combination of standard solutions: pH 7 and one of the others Manual calibration: Freely selectable, but the difference should be over 2 pH. Temperature calibration (one point). | ||
Transmission output hold feature |
| ||
Self-diagnosis function |
| ||
Temperature compensation element | Selectable from compensation ON (350Ω (25℃), 500Ω (25℃),1 kΩ (0℃)6.8 kΩ (25℃) | ||
Temperature- compensated range | 0℃ to 100℃ | ||
Ambient temperature | −5℃ to 45℃ | ||
Relative humidity | 20% to 85% (without dew condensation) | ||
Storage temperature | −25℃ to 65℃ | ||
Power supply | Rated voltage 100 V AC to 240 V AC, 50 Hz/60 Hz, 10 VA (max.) | ||
Structure | Indoor-use panel installation type Panel case: ABS, Terminal: PBT Panel: water proof structure | ||
Protective structure | Panel: IP65 (IEC60529, JIS C0920) Rear case: IP20, Terminal: IP00 Class II device (IEC61010-1) Pollution level 2 (IEC61010-1) | ||
Conforming standards | CE Marking | EMC Directive: EN61326-1 Low Voltage Directive: EN61010-1 RoHS Directive: EN50581 | Immunity (Industrial electromagnetic environment) Noise increase ±0.2 pH or less |
FCC Rule | FCC Part15 | ||
External dimensions | 48 (W) × 96 (H) × 115 (D) mm Case depth: approx. 105 mm | ||
Mass | Approx. 400 g |
Do you have any questions or requests? Use this form to contact our specialists. * These fields are mandatory.
Automated Organic Analysis & Early Warning Sentinel
For Instant Water Quality Reports
Standard type
For measurement of low-conductivity water and non-aqueous solvents.
Carbon Sensor Conductivity Meter (Low concentration type)
Industrial resistivity meter