MEXA-ONE-FT

MEXA-ONE-FT - FTIR Motor Exhaust Gas Analyzer

FTIR Motor Exhaust Gas Analyzer

The MEXA-ONE-FT Series simultaneously measures concentrations of multiple components, including NH3, N2O, CH4, CO and CO2, in undiluted exhaust gas using the Fourier transform infrared (FTIR) method in combination with a multivariate analysis algorithm. It incorporates HORIBA's proprietary compact interferometer which yields superior performance and high accuracy.
The system complies with NH3 testing requirements of Euro VI, the EU regulation for heavy duty emissions testing.
The MEXA-ONE-FT can be calibrated according to customer requirements with up to 6 component sets to cover a wide range of applications.
FTIR spectroscopy allows the direct measurement of NO2 emissions and the independent quantification of individual hydrocarbon species. This makes the MEXA-ONE-FT system an ideal solution for detailed research and development tasks.
Due to the measurement principle, the time-consuming daily span gas calibration is omitted.
In addition, usability and flexibility are realized by “HORIBA ONE PLATFORM” which enables to extend the system with other ONE analyzers to a complete solution suitable for both certification and R&D purposes.

Segment: Automotive
Manufacturing Company: HORIBA, Ltd.

Applications

  • Euro VI compliance "MEXA-ONE-FT-E"

NH3 response time (t10-90): 5 seconds or less
Sampling temperature: 191°C

  • Simultaneous and continuous measurement of up to 28 components such as NO, NO2, N2O, NH3

・R&D for alcohol-based fuels
・Development of exhaust after-treatment systems
・Measurement of greenhouse gas

 

  • Tools for certification tests and R&D

Combined with the MEXA-ONE series, which is used for emission certification test, the MEXA-ONE-FT simultaneously measures NH3 along with various regulated and unregulated exhaust gas components. Options for THC and O2 measurements are available.

Outline

ModelStandard type: MEXA-ONE-FT
Euro VI compliance type: MEXA-ONE-FT-E
Conformed standardsCE, FCC, KC
Target components and ranges*1

Maximum 28 components 
Standard component set: for gasoline/diesel vehicle (18 components and “NO+NO2”)
Optional component set: for CNG-fueled vehicle (16 components and “NO+NO2”)

ApplicationsEmissions performance tests of various engines, evaluation of after-treatment systems, R&D for alcohol-based fuels, Certification of NH3emissions according to Euro VI
PrincipleFourier transform infrared spectroscopy (FTIR)
Outputs/InputsMonitoring on 17-inch LCD
Sample line temperature

MEXA-ONE-FT: 113 ± 6ºC
MEXA-ONE-FT-E: 191 ± 6ºC

Sample gas flow rateMEXA-ONE-FT: 4.0 ± 0.5 L/min
MEXA-ONE-FT-E: 14 ± 1 L/min
Operating environment

Ambient temperature: 15 to 35ºC 
Ambient humidity: less than 80% relative humidity

 System Configuration 

Liquid nitrogenReplenishing amount: approx. 500 ml (including evaporation)
Duration: 10 hours or more (The dewar vacuum must be renewed annually)
Power supply voltageAny of the following power line voltages and frequencies (to be specified when ordering)
・ 100 ± 10 V AC, 50/60 ± 1 Hz, single phase
・ 110 ± 11 V AC, 50/60 ± 1 Hz, single phase
・ 115 ± 11.5 V AC, 50/60 ± 1 Hz, single phase
・ 120 ± 12 V AC, 60 ± 1 Hz, single phase
・ 200 ± 20 V AC, 50/60 ± 1 Hz, single phase
・ 220 ± 22 V AC, 50 ± 1 Hz, single phase
・ 230 (207~250) V AC, 50 ± 1 Hz, single phase
・ 240 (216~250) V AC, 50 ± 1 Hz, single phase
Power requirementsApprox. 4.0 kVA as maximum, approx. 3.0 kVA at stable state
Dimensions

655 (W) x 855 (D) x 1970 (H) mm
(excluding protruding objects)

MassApprox. 460 kg

Performance

Zero noise*2

Less than 1.0% of full scale
(depending on measurement components)

Response time*3
(t10-90)
MEXA-ONE-FT: 5 seconds or less when
switching from N2 to 200 ppm CO.
MEXA-ONE-FT-E: 5 seconds or less
when switching from N2 to 200 ppm NH3
Repeatability *4 *5Within ±1.0% of full scale
Zero drift/span drift*4 *5Within ±1.0% of full scale/8 hours
Warm-up timeLess than 2 hours after turning on
the main power

Options

Sample line selector8 lines
Analog output board

Analog output: 0-10V Max. 16 channels
(Output components are selectable)
A connection cable (Max. 15m)

Host communication cableLAN cable
NH3 filter elementSilica fiber (φ19.6 mm x L90 mm)
THC measurementPrinciple: Heated flame ionization detection (HFID)
Measurement range: from 0-10 ppmC to 0-60000 ppmC
O2 measurement

Principle: Magnetopneumatic Detector (MPD) Measurement range: from 0-1 vol% to 0-25 vol%

Automatic replenishing device for liquid nitrogen

Maximum storage amount: Approx. 12 L 
Duration: 120 hours or more
(depending on the vacuum degree of dewar)
For more details, please contact HORIBA

*1: To be specified when ordering.
*2: The zero noise is defined as the twofold standard deviation of the concentration reading.
*3: Using a sample probe (0.3 m), a heated filter and a heated pipe (6 m) without contamination of the sample line.
*4: Confirmed with 200 ppm CO, 500 ppm CH4 and 200 ppm C2H6.
*5: Span gas with more than 80% of the full scale concentration shall be used.

Engine testing in automobiles and components
Engine testing in automobiles and components
Catalyst testing in automobiles and components
Catalyst testing in automobiles and components

제품 문의

HORIBA제품의 자세한 정보를 원하시면, 아래의 양식에 내용을 입력을 부탁드립니다.

* 는 필수입력항목입니다.

Related products

CVS-ONE
CVS-ONE

희석 배기가스 분석기

EXFM-ONE
EXFM-ONE

초음파 배기 유량계

FTIR 배기 가스 분석기 FTX-ONE-CS/RS
FTIR 배기 가스 분석기 FTX-ONE-CS/RS

FTIR Exhaust Gas Analyzer Direct Measurement Type

FTX-ONE-CL
FTX-ONE-CL

FTIR 배기 가스 분석기 희석 측정 타입

MDLT-ONE
MDLT-ONE

부분 희석 PM 샘플 장치

MEXA-2000SPCS series
MEXA-2000SPCS series

배기가스 고체 입자 카운팅 시스템

MEXA-ONE
MEXA-ONE

배기 가스 측정 장치

MEXA-ONE-QL-NX
MEXA-ONE-QL-NX

레이저 분광 배기가스 측정 장치

51 series
51 series

방폭형 가스 분석기

AP-370 series
AP-370 series

Air Pollution Monitor

AQMS
AQMS

Ambient Air Quality Monitoring System

CEMS
CEMS

Continuous Emission Monitoring System

ENDA-5000 Series
ENDA-5000 Series

굴뚝 배출가스 분석장치

FTIR 배기 가스 분석기 FTX-ONE-CS/RS
FTIR 배기 가스 분석기 FTX-ONE-CS/RS

FTIR Exhaust Gas Analyzer Direct Measurement Type

FTX-ONE-CL
FTX-ONE-CL

FTIR 배기 가스 분석기 희석 측정 타입

IR-200
IR-200

Gas Monitor for Chamber Cleaning End Point Monitoring

IR-400
IR-400

High-grade type Gas Monitor for Chamber Cleaning End Point Monitoring

MEXA-ONE
MEXA-ONE

배기 가스 측정 장치

MEXA-ONE-QL-NX
MEXA-ONE-QL-NX

레이저 분광 배기가스 측정 장치

NZ-3000
NZ-3000

지르코니아 산소 분석기

OBS-ONE GS Unit
OBS-ONE GS Unit

차량 탑재형 배기가스 측정 장치

OBS-ONE IRLAM Unit
OBS-ONE IRLAM Unit

NH3/N2O 대응용 차량 탑재형 배기가스 측정 장치

PG-300 시리즈
PG-300 시리즈

휴대용 가스분석기

Process
Process

Process Analyzer System

PX-375
PX-375

PM & 중금속 분석기

VA-5000
VA-5000

범용 멀티 가스 분석기

FTIR 배기 가스 분석기 FTX-ONE-CS/RS
FTIR 배기 가스 분석기 FTX-ONE-CS/RS

FTIR Exhaust Gas Analyzer Direct Measurement Type

FTX-ONE-CL
FTX-ONE-CL

FTIR 배기 가스 분석기 희석 측정 타입