| Product name | Continuous Particulate Monitor with X-ray Fluorescence |
|---|---|
| Model | PX-375 |
| Measured object | Particulate matter (PM10, PM2.5, TSP) |
| Measurement content | Particulate mass concentration and element concentration |
Common
| Flow rate | 16.7 L/min |
|---|---|
| Sampling pump | Linear drive system, externally installed |
| Filter tape | None-woven PTFE fabric filter |
| Spot tape interval | 20 / 25 / 50 / 100mm selectable |
| Filter tape replacement interval | Approx. 5 months (In case of 20 mm spot interval, 1 hour sampling time) |
| Ambient operation temperature | 10˚C~30˚C |
| Relative humidity | 0~80% RH noncondensing |
| Altitude | 1000m or less |
| Power supply | AC100V~240V ±10%, 50/60Hz±1% |
| Power consumption | Approx. 400VA |
| External dimension | 430mm(W)×550mm(D)×285mm(H) (without sampling pipe and measurement head) |
| Weight | Approx. 49kg |
| Data output | CSV file (Average PM mass and elemental concentration) |
| External connection | EthernetTM, USB, RS-232C* (option) |
*Please consult about communication and instrument composition separately.
Mass analyzer unit
| Measurement method | Beta-ray attenuation |
|---|---|
| PM10 | US EPA Louvered PM10 Inlet |
| PM2.5 | BGI VSCCTM Cyclone |
| TSP | TSP Inlet |
| Measurement range | 0~200 / 500 / 1000 μg/m3 |
| Repeatability | ±2% (against reference foil value) |
| Span drift | ±3 % (24h) |
| Lowest detection limit (2σ) | ±4 μg/m3 (24h) |
| Sampling and measurement cycle | 0.5 / 1 / 2 / 3 / 4 / 6 / 8 / 12 / 24 hours |
Element analyzer unit
| Measurement method | Energy dispersive X-ray fluorescence spectrometry |
|---|---|
| Detectable elements | See Table 1 "Detectable Elements". 32 standard parameters: Al, Si, S, K, Ca, Ti, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, As, Se, Rb, Sr, Zr, Pb, Ga, Ge, Y, Pd, Ag, In, Sn, Sb, Te, Cs, Ce, Bi |
| Primary X-ray filter | Automatic switching for light metals/heavy metals |
| Tube voltage | Automatic switching for 15kV/50kV |
| Detector | SDD (Silicon Drift Detector) |
| Sample image | CMOS camera |
| Lowest detection limit (2σ) | See Table 2 "Lowest Detection Limit (Example)" |
| Measurement range | depends on measurement time |
| Analysis time | 100 / 200 / 500 / 1000 / 2000 / 5000 / 10000s selectable |
| Calibration material for X-ray intensity for standard parameter | NIST SRM 2783, other materials (option) |
| Safety functions for X-ray | Internal lock system Key switch X-ray indication light |
Black Particles
| Measurement method | Energy dispersive X-ray fluorescence spectrometry |
|---|---|
| Measured object | Particulate matter of black color |
| Detector | CMOS camera |
| Light source | LED |
| BP measurement range | 0-200 |
| Lowest Detection Limit | <5 BP value |
| Span drift | BP value: ±2% (24 h) |
| Spot tape interval | 25 mm only |
| Sampling cycle | Follow the setting of mass and element measurement |
| Repeatability | ±2% of full scale (Typical), ±3% of full scale (Max) |
| Calibration material | Neutral density filter |