OCMA-550

Oil concentration meter

Automatic measurement with a single switch.

One-touch operation with a single button press allows for ease of use by anybody. Can be used for a wide range of applications such as wastewater monitoring and component quality control.

OIL

Division: Water Pollution
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Quick and easy measurement in approximately 3 minutes*1

Measurements can be made easily and quickly by simply pressing the start button. Compared to the n-hexane method, measurement times can be significantly reduced.
*Excluding time for activities such as warming, calibration, rinsing

Capable of measuring low-boiling point oil components

The n-hexane method requires the evaporation of the oil extraction solvent, which leads to volatilization of low boiling point oils (such as gasoline). The OCMA-500 series eliminates the need for an evaporation process, allowing for the prevention of volatilization in low-boiling point oil components.

Oil content can be displayed as a concentration (value)

Unlike the n-hexane method, which calculates concentration by measuring the weight of oil content, OCMA-550 can directly display oil concentration.
*Please contact us for inquiries regarding the types of oil that can be detected.

Environmental-friendly, safe, and cost-effective oil extraction solvents are in use

The domestic model OCMA-505-H/555-H uses solvent H-519 for oil extraction, while the international model OCMA-500/550 uses solvent S-316. Both H-519 and S-316 are not regulated under the Montreal Protocol (Ozone Layer Protection Act) and the PRTR system (Act on the Evaluation of Chemical Substances and Regulation of Their Manufacture, etc. (CSCL)). Furthermore, compared to the traditional solvent H-997, solvent H-519 has achieved similar regeneration efficiency to solvent S-316, allowing for a reduction in running costs.

ModelOCMA-550
Product nameOil content analyzer
Measurement methodSolvent extraction - non-dispersive infrared absorption analysis method
Measured objectsSubstances extracted from sample water into solvent and having infrared absorption near a wavelength from 3.4 μm to 3.5 μm
Measurement range0 mg/L to 200 mg/L
ResolutionFor mg/L 0 to 99.9: 0.1, 100 to 200: 1
For mg/g, mg/kg, mg/PC 0 to 9.99: 0.01, 10.0 to 99.9: 0.1, 100 to 200: 1
Abs.: 0 to 1.000 Abs.: 0.001 Abs.
Repeatability0 mg/L to 9.9 mg/L: ±0.4 mg/L ±1 dig.
10.0 mg/L to 99.9 mg/L: ±2.0 mg/L ±1 dig.
100 mg/L to 200 mg/L: ±4 mg/L ±1 dig. * For standard liquids
Display method3.5 inches, 320×240 dots
Backlight Color graphic LCD
Calibration methodSelect each optionally zero calibration and span calibration
Extraction solventS-316
Amount of extraction solvent requiredApprox. 6.5 mL
Extraction methodUsing the extraction solvent, and extracted manually outside the product
Ambient operating temperature0°C to 40°C (no condensation)
Power supply100 V to 240 V AC ±10%, 50/60 Hz
Power consumption100 V to 240 V AC, Approx. 60 VA
External dimensions195 (H) × 253 (W) × 293 (D) mm
MassApprox. 5 kg
External outputOutput to a USB memory stick
Functions
  • 300-item data memory (measurement history)
  • Stabilized measurement value display
  • Self error determination
  • Clock
  • Conversion of measurement units
  • Compliant with ASTM D7066-4