Field-installation Type Electric Conductivity Meter (Two-Wire Transmission, For Low-concentration)
HE-300C connects conductivity sensor (ESH/FS Series) and measures conductivity and temperature in the sample water. It is suitable for monitoring pure water/boiler water drainage recovery control and inline measurement of pharmaceutical water conductivity according to USP<645> quality guidelines.
Conductivity: 0 - 2000 μs/cm, 0 -200.0 ms/cm
(TDS conversion 0 - 2000 mg/L)
Temperature: 0 - 100 ℃
Product name | Conductivity meter | ||||
---|---|---|---|---|---|
Model | HE-300C | ||||
Sensors in combination | 2-electrode method conductivity sensor ESH and FS Series sensor of cell constant 0.01/cm, 0.1/cm, or 1.0/cm | ||||
Measurable range (*Note 1) | Cell constant (/cm) | 0.01 (ESH-001) | 0.1 (ESH/FS-01) | 1.0 (ESH-1) | |
Unit | |||||
Conductivity | μS/cm | 0.000 to 2.000 0.00 to 10.00 | 0.000 to 2.000 0.00 to 20.00 0.0 to 100.0 | 0.0 to 200.0 0 to 1000 | |
mS/m | 0.0000 to 0.2000 0.000 to 1.000 | 0.0000 to 0.2000 0.000 to 2.000 0.00 to 10.00 | 0.00 to 20.00 0.0 to 100.0 | ||
TDS conversion | mg/L | 0.00 to 2.00 0.0 to 10.0 | 0.00 to 2.00 0.0 to 20.0 0 to 100 | 0 to 200 0 to 1000 | |
Temperature | ℃ | 0 to 100 (display range: −10℃ to 160℃) | |||
Display resolution | Conductivity / TDS conversion | As shown in the measuring table. | |||
Temperature | 0.01℃ | ||||
Performance | Conductivity | Repeatability | Within ±0.5% of the full scale (response for equivalent input) Within ±5% of the full scale in the range of 2000 μS/cm or 200 mS/m for FS-01 sensor | ||
Linearity | Within ±0.5% of the full scale (response for equivalent input) Within ±5% of the full scale in the range of 2000 μS/cm or 200 mS/m for FS-01 sensor | ||||
TDS conversion | Repeatability | Within ±1.5% of full-scale value (response for equivalent input) | |||
Linearity | Within ±1.5% of full-scale value (response for equivalent input) | ||||
Temperature | Repeatability | ±0.1℃(response for equivalent input) | |||
Linearity | ±0.5℃(response for equivalent input) | ||||
Transmission output | Output type | 4 mA to 20 mA DC input/output insulation type (two-wire transmission type) | |||
Load resistance | Maximum: 600 Ω. Case of 24 V DC power supply (*Note 3) | ||||
Linearity | Within ±0.08 mA (output only) | ||||
Repeatability | Within ±0.02 mA (output only) | ||||
Output range | Conductivity: Free setting within a measuring range. | ||||
Occasional out for error | Hold or burnout to either 3.8 mA or 21 mA | ||||
Transmission hold | In the maintenance mode, transmission signal is held at the latest value or preset value. In the calibration mode, transmission signal can be alive or held. | ||||
Contact input | Number of input points | 1 | |||
Contact type | No-voltage a-contact | ||||
Conditions | ON resistance: 40 Ω max. Open voltage: 1.2 VDC Short-circuit current: 21 mA DC max. | ||||
Contact function | External input for transmission holding. | ||||
Temperature compensation | Temperature element | Platinum resistor: 1 kΩ (0℃) | |||
Temperature compensation |
| ||||
Temperature compensation range | 0℃ to 100℃ (However, the compensation calculation is extended lower than 0℃ or higher than 100℃.) | ||||
Temperature calibration | 1 point calibration comparing to reference thermometer | ||||
Calibration | Conductivity | Based on parameter input of coefficient for the sensor cell constant. | |||
TDS conversion | Conversion by user-defined coefficient value (TDS calculation is 0.30 to 1.00 times of sample conductivity by μS/cm.) | ||||
Self-check | Sensor diagnosis error | Temperature sensor short-circuit, temperature sensor disconnection, and out of the temperature measurement range | |||
Meter error | CPU error, ADC error, and memory error | ||||
Operating temperature range | −20℃ to 60℃ (without freeze) | ||||
Operating humidity range | Relative humidity: 5% to 90% (without condensation) | ||||
Storage temperature | −25℃ to 65℃ | ||||
Power supply | Rated voltage | 24 V DC (operating voltage range: 21 V to 32 V DC) (*Note 3) | |||
Power consumption | 0.6 W max. | ||||
Compatible standards | CE marking | EMC directive: EN61326-1 RoHS Directive: EN50581 | |||
Immunity Industrial electromagnetic environment | Electostatic discharge | IEC61000-4-2 | |||
Electromagnetic field of radiated radio frequency | IEC61000-4-3 | ||||
Electric fast transient/burst | IEC61000-4-4 | ||||
Surge | IEC61000-4-5 (*Note 2) | ||||
Conducted interference induced by radio frequency | IEC61000-4-6 | ||||
Emission Class A | Radiated disturbance | CICPR 11 CLASS A | |||
Noise terminal voltage | CISPR 11 CLASS A | ||||
FCC Rules | Part15 CLASS A | ||||
Structure | Installation | Outdoor installation type | |||
Installation method | 50 Mount on 50 A pole or wall | ||||
Protection code | IP65 | ||||
Case material | Aluminum alloy (coated with epoxy-denatured melamine resin) | ||||
Material of fittings | SUS304 | ||||
Material of hood | SUS 304 stainless steel (coated with epoxy-denatured melamine resin) | ||||
Material of window | Polycarbonate | ||||
Display element | Reflective monochrome LCD | ||||
External dimensions | 180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets) | ||||
Mass | Main body: Approx. 2.8 kg; cover and mounting brackets: Approx. 1 kg |
Do you have any questions or requests? Use this form to contact our specialists. * These fields are mandatory.
For Instant Water Quality Reports
Carbon Sensor Conductivity Meter (Low concentration type)
Industrial resistivity meter
KOH Monitor
Wide range TMAH Concentration Monitor
High Precision TMAH Concentration Monitor
Flat Carbon Sensor Conductivity meter
Carbon Sensor Resistivity Meter
2-Channel Resistivity Meter