| Model | HE-960R-GC | |||||
|---|---|---|---|---|---|---|
| Measurement method | Electrode type (2-electrode method) | |||||
| Cell constant | Approx. 0.1/cm | |||||
| Measuring range | Measurement unit | kΩ·m | 0 to 2.00 | 0 to 20.0 | 0 to 200.0 | 0 to 1000 * |
| MΩ·cm | 0 to 0.200 | 0 to 2.00 | 0 to 20.00 | 0 to 100.0 * | ||
| *: Measurable without temperature compensation Temperature: 0°C to 100°C (Select your desired decimal point from 0, 1, and 2 digits.) | ||||||
| Repeatability Within | ±0.5% of the full scale (in equivalent input) | |||||
| Linearity Within | ±0.5% of the full scale (in equivalent input) | |||||
| Transmission output | 4 mA to 20 mA DC: input/output isolated type Maximum load resistance: 900Ω Transmission output range: Freely selectable within the measurement range | |||||
| Contact output | Outputs: 2 points Alarm contact output (R1 and R2) | |||||
| Calibration function | Specific resistance Temperature | |||||
| Transmission output hold feature | · Previous value hold · Optional value hold · Continuous Selectable from the above (However, only the previous value hold is available in the maintenance Mode.) | |||||
| Self-diagnosis function | · Sensor diagnosis · Out of the measurement range · A/D converter scale over · Converter error | |||||
| Temperature compensation | · Based on the temperature characteristics of extra deionized water (reference temperature:25°C) · Based on the reference temperature and user-defined temperature coefficient (reference temperature: 5°C to 95°C) · No temperature compensation | |||||
| Temperature - compensated range | 0°C to 100°C | |||||
| Extra deionized water specific resistance selection | Measurement unit | kΩ·m | 182.3 (standard), 181.8, 182.4 | Select from options shown on the left. | ||
| MΩ·cm | 18.23 (standard), 18.18, 18.24 | |||||
| Clipping function | When the measured value is above the upper limit of the measurement range derived from the specified specific resistance, the specified resistance is used as the measured value. | |||||
| Temperature compensation element | Platinum resistance 1000Ω/0°C | |||||
| Power supply | Rated voltage 24V DC 5W(max) | |||||
| Structure | Indoor-use panel installation type Panel: IP65 dust and water proof structure | |||||
| Protective structure | Panel: IP65 (IEC60529, JIS C0920) | |||||
| Conforming standards | CE Marking FCC Rule | |||||
| External dimensions | 96(W) mm x 96 (H) mm x 115 (D) mm Case depth: approx. 105 mm (when panel-mounted) | |||||
| Mass | Approx. 500 g | |||||
| Compatible sensors | ERF-series specific resistance GC sensor, cell constant 0.1/cm | |||||
외형도
![]()