Experience HORIBA, the Global Single Source Solutions Provider with over 50 years of technological advancements and unparalleled reliability, dependability and operability.The GI-700 stack gas analyzer is designed to continuously measure the concentrations of nitrogen oxides (NOx), sulfur dioxide (SO2), carbon monoxide (CO), carbon dioxide (CO2), oxygen (O2), and ammonia (NH3) in stack flue gas streams.
Model | GI-700 | ||||||
Component | NOx | SO2 | CO | CO2 | O2 | NH3*3 | |
Measurement methods*1 | NDIR | NDIR | NDIR | NDIR | MPA | CLA | |
Range*1 | Standard | 10~5000 ppm | 10~5000 ppm | 50~5000 ppm | 5~25 vol% | 10~25 vol% | 10-100 |
| Optional | 5 ppm~ | 5 ppm~ | 5 ppm~ | 10 vol%~ | - | - |
Range Ratio*1 | Standard | 1:10 | 1:10 | 1:10 | 1:5 | 1:5 | 1:10 |
Optional | 1:20 | 1:20 | 1:20*2 | 1:5 | - | - | |
Indicated error | ± 1.0 % of full-scale value for each component and range (Standard range ratio) ± 2.0 % of full-scale value for each component and range (Optinonal range ratio) | ||||||
Zero drift | ± 1.0%/week of full scale (standard range) (if the ambient temperature change is ≤ 5°C) | ||||||
Span drift | ± 2.0%/week of full scale (if the ambient temperature change is ≤ 5°C) | ||||||
Response time | T90=45s from the analyzer inlet | ||||||
Interference | ±2.0% of full scale | ||||||
Environment | Temperature | 0° to 40°C (No direct sun light or no radiation heat) | |||||
Humidity | ≤ 90% | ||||||
Vibration | 100Hz, 0.3m/s2 or less | ||||||
Dust | Less than environment standard | ||||||
Measuring Gas Condition | Temperature | Ambient temperature | |||||
Dust | None | ||||||
Standard gas composition | NO :Up to 5000 ppm | ||||||
Sample inlet tube | Polytetrafluoroethylene tube (6 mm O.D./4 mm I.D.) | ||||||
Sample gas pressure | Atmospheric pressure ± 4 in. W.C. | ||||||
Pressure control | Not including (Vacuum regulator (-9.8 kPa)) | ||||||
Output(Option) | Analog output: 6 channels 4-20mA, 0-1V, 0-5V | ||||||
Calibration gas | Zero gas : Nitrogen | ||||||
Carrier gas for (Oxygen analyzer | Nitrogen (for N2 carrier specification), Air (for air carrier specification) | ||||||
Air for ozone generation | Dry air (saturated at ≤ -30°C) | ||||||
Power supply | 100 - 240 VAC (Depends on components) | ||||||
Power frequency | 50/ 60 Hz for common use | ||||||
Power consumption | 240 VA | ||||||
Exterior dimensions/Mass | 5U below in (H)19 in. (W)22 in. (D) / Approx. 19kg (depends on the incorporated components) | ||||||
Materials in contract with sample gas | PTFE, Nylon, PP, FKM, SS304, SS316 | ||||||
Serial Communications | ModbusTM, Ethernet (Modbus address will be disclosed) *Modbus is a trademark of Schneider Electric Inc. |
*1: Depends on measurement range specifications.
*2: A range ratio of 1: 40 is also possible for special applications.
*3: A range ratio of CLA NH3 and CLA NOx should be 1: 5 or less.
Do you have any questions or requests? Use this form to contact our specialists. * These fields are mandatory.
Stack Gas Analysis System
Portable Gas Analyzer
Multi-Component Gas Analyzer
Air Pollution Monitor
Ambient Ammonium Monitor
Ambient Hydrogen Sulfide Monitor
Stack Gas Analysis System
FTIR Exhaust Gas Analyzer Dilute Measurement Type
FTIR Exhaust Gas Analyzer Direct Measurement Type
Heated Type NOx Analyzer
Heated Type THC Analyzer
Motor Exhaust Gas Sulfur Analyzer
Motor Exhaust Gas Analyzer
FTIR Motor Exhaust Gas Analyzer
Laser Spectroscopic Motor Exhaust Gas Analyzer
On-board Emissions Measurement System
On-board Emissions Measurement System for NH3/N2O
Portable Gas Analyzer
Multi-Component Gas Analyzer
Air Pollution Monitor
Ambient Ammonium Monitor
Oxygen/Nitrogen/Hydrogen Analyzers
Carbon/Sulfur Analyzer
Stack Gas Analysis System
Heated Type NOx Analyzer
Motor Exhaust Gas Analyzer
On-board Emissions Measurement System
On-board Emissions Measurement System for NH3/N2O
Portable Gas Analyzer
Multi-Component Gas Analyzer