VA-3000/VS-3000 Multi-component Analyzer

Descripción

From environmental monitoring to developing new energy sources and chemicals for the new era, gas analysis systems are faced with needs and challenges that have changed dramatically over time. Responding to these new needs, HORIBA has developed the VA-3000, the versatile gas analyzer that's ready for the future. A single analyzer is now capable of measuring a wider selection of gas components utilizing many different types of sensor technology. Non-dispersive infrared (NDIR) modules are available to measure gases such as CO, CO2, NO, N2O, CH4, SO2 and others. A chemiluminescent sensor module may be included to measure NO or NOx with a standard converter. And four different types of sensor modules are available to measure O2 - galvanic, paramagnetic, zirconium oxide or magnetopneumatic. Designed to be compact and lightweight, the VA-3000 may have up to three sensor modules installed in a single analyzer case. This unique feature means the VA-3000 can be used today for a broad range of applications including research and development or environmental pollution monitoring, where efficiency and space-saving are crucial.

VA-3000 Series product literature

Características

  • Use to measure the concentration of many different gases
    A single VA/VS-3000 analyzer is capable of simultaneously measuring multiple gas components such as CO, CO2, O2, NOx, SO2, CH4, N2O and other gases. With the VS-3000 sampling system, the VA/VS-3000 will meet the needs of a diverse range of applications.
    CO, CO2, NO, NOx, SO2, CH4, C2H4, N2O, O2 and others.
  • Many combinations of sensor modules are possible
    Development of smaller sensor modules now makes it possible to install up to three sensor modules in a single unit. You can install three NDIR modules in a single case or substitute a chemiluminescent or oxygen module for any of the NDIR modules.
  • Wide selection of ranges and flexibility for many applications
    The VA-3000 also features dynamic ranges up to 20x the base (first) range. And the dynamic range can be further expanded by installing multiple sensor modules for the same component gas in a single analyzer case. Now a single unit can perform measurements over a wide range from ppm to 100 percent concentrations. With our diversified lineup of modules, you can always choose the optimum range for your application.
  • High precision assured
    The VA-3000 achieves repeatability of ±0.5% of F.S.R. (Full Scale Range), linearity of ±1.0% F.S.R., and drift of less than ±2.0% per week without requiring recalibration with a gas standard saving you money. For optimum precision, use the auto daily calibration feature, to virtually eliminate drift. Or specify the optional gas filled calibration cell that is used to routinely check and adjust the calibration of the NDIR modules without the need for external gas cylinders. With these features, high precision measurement is maintained throughout the entire measurement range.
  • Standard RS-232C interface makes data collection easy and remote operation possible
    Analog outputs are also available for direct connection to control systems.
  • HORIBA provides reliable gas analysis systems
    As an expert on gas analysis systems, with a wealth of experience, HORIBA can provide consulting services and/or appropriate hardware for pretreatment of even the most difficult sample gases.

Producido por HORIBA

Especificaciones

Analyzer Unit

Specification

Representative measuring object

NDIR : CO, CO2, N2O, CH4, SO2
CLD : NOx
MPA, Galvanic, Zirconia, Paramagnetic : O2

Power supply

AC 100 to 120 V or AC 200 to 240 V, 50/60Hz, (Voltage should be specified)

Power consumption

Approx. 200VA for one NDIR module to 300VA for three modules

Measuring gas condition

Flow rate Approx. 0.5L/min

Sample condition

Temperature: Ambient temperature, Water: Saturated at 5°C or less, Pressure: 490kPa, No dust, Corrosive, flammable, and explosive gases are not allowed

Gas connections

1/4 F NPT except exhaust is 1/4 tube; can be optionally installed on front panel; a single gas inlet is provided standard; the gas flows sequentially from one module to the next; as an option separate gas inlets to be provided for each module.

Gas Tubing

Teflon; 316 stainless steel optional

Environment conditions

Temperature: 0 to 40°C, Relative humidity: 90% or less

Input/Output

Analog outputs

DC 0 to 1 V (Non-isolated output), or DC 4 to 20 mA (Non-isolated output); Two per modules

Serial communication

Two RS-232C ports; additional RS-485 port And MODBUS or Ethernet Protocols are optional

Display

LCD 118.18 mm (W) x 89.38 (H), 320 dots (W) x 240 dots (H)

Case

19-inch panel mount (Standard)

Exterior dimensions

430 mm (W) x 132 mm (H) x 550 mm (D)

Mass

Approx. 20kg (Approx. 44 lbs) for three modules in one analyzer unit

 

Performance

NDIR

Standard range*                CO        Min. 0-200ppm, Max. 0-100vol%

                                          CO2       Min. 0-100ppm, Max. 0-100vol%

                                          N2O       Min. 0-100ppm, Max. 0-5000ppm

                                          CH4       Min. 0-200ppm, 0-100vol%

                                          SO2      Min. 0-200ppm, 0-10vol%

Repeatability                      ±0.5% of full scale

Linearity                             ±1.0% of full scale (Range ratio 10 times or less)

Drift (Zero & Span)             ±2.0% of full scale per week                         

Response time                   Within 30 second (T90), Within 45 second (T95)

Warm-up time                    Approx. 20 minutes

 *4 ranges with a maximum ratio between the lowest and highest range (range ratio) of 10x ( or 20x as an option, although this may be limited by the cell length).

 

CLA

Standard range*                NOx        Min. 0-20ppm, Max. 0-5000ppm

Repeatability                      ±0.5% of full scale (±1.0% for 0-100ppm)

Linearity                             ±1.0% of full scale

Drift (Zero & Span)             ±2.0% of full scale per week                         

Response time                   Within 30 second (T90), Within 45 second (T95)

Warm-up time                    Approx. 60 minutes

 *8 ranges with a maximum ratio of 100x (if the maximum range exceeds 2000ppm, the minimum range should be 5ppm or more).

 

Magneto pneumatic

Standard range*                O2         Min. 0-5vol%, Mix. 0-100vol%

Repeatability                      ±0.5% of full scale

Linearity                             ±1.0% of full scale

Drift (Zero & Span)             ±2.0% of full scale per week                         

Response time                   Within 30 second (T90), Within 45 second (T95)

Warm-up time                    Approx. 60 minutes

 *4 ranges with a maximum range ratio of 10x.

 

Galvanic

Standard range*                O2         Min. 0-5vol%, Max. 0-25vol%

Repeatability                      ±0.5% of full scale

Linearity                             ±1.0% of full scale

Drift (Zero & Span)             ±1.0% of full scale per day                           

Response time                   Within 30 second (T90), Within 45 second (T95)

Warm-up time                    Approx. 20 minutes

 *4 ranges with a maximum range ratio of 5x.

 

Zirconia

Standard range*                O2         Min. 0-5vol%, Max. 0-25vol%

Repeatability                      ±0.5% of full scale

Linearity                             ±1.0% of full scale

Drift (Zero & Span)             ±2.0% of full scale per week                         

Response time                   Within 30 second (T90), Within 45 second (T95)

Warm-up time                    Approx. 40 minutes

 *4 ranges with a maximum range ratio of 5x.

 

Paramagnetic

Standard range*                O2         Min. 0-5vol%, Max. 0-25vol%

Repeatability                      ±0.5% of full scale

Linearity                             ±1.0% of full scale

Drift (Zero & Span)             ±1.0% of full scale per week                         

Response time                   Within 30 second (T90), Within 45 second (T95)

Warm-up time                    Approx. 20 minutes

 *4 ranges with a maximum range ratio of 5x.

 

Sampling Unit

Specification

Application    VS-3001 for NDIR, Galvanic, Zirconia and Paramagnetic

                     VS-3002 for NDIR, Galvanic, Zirconia, MPA and Paramagnetic

                     VS-3003 for NDIR, Galvanic, Zirconia, Paramagnetic and CLD

                     VS-3004 for NDIR, Galvanic, Zirconia, Paramagnetic, MPA and CLD

Form        19-inch panel mount

Sampling method      5 degrees C sampling

Power supply    100 - 120V AC or 200 – 240V AC, 50/60Hz (specify voltage)

Dimensions        430 mm (W) x 221 mm (H) x 550 mm (D)

Mass         VS-3001 approx. 14kg

                 VS-3002 approx. 16kg

                 VS-3003 approx. 19kg

                 VS-3004 approx. 20kg