Organic Pollutant Monitor
The OPSA-150 is a organic pollutant monitor that uses HORIBA's proprietary Rotary Cell Length Modulation, a measuring technique incorporating 25 years of expertise. The unit can be used as an organic pollutant monitor at drainage systems for determining compliance with COD monitoring regulations, for monitoring quality of water measuring levels of organic matter at water supply intakes, and as an organic monitor on process lines (phenol meter).
*Pole mount, outdoor cover, and analysis panel are available as options.
(JAPAN Patent No. 4627022, KOREA Patent No. 10-0840034, CHINA Patent No. ZL200610110930.8)
Zero drift correction is perfomed in every cycle of measurement. This correction can eliminate the effect of any interference on cells.
Cell length modulation provides readings from multiple measurements, making available results from various cell lengths. This allows the single device to make measurements on a large range of concentrations from as low as 0 to 0.1 Abs, up to 0 to 5.0 Abs.
* CompactFlash is a registered trademark or trademark of SanDisc Corporation in the United States and other countries.
Control of the OPSA-150 is operated via the intuitively designed touch panel. An interactive setup screen eliminates any doubts that a user may have while operating the unit. The Converted COD and TURB Display allow readings to be confirmed on site.
Internal data memory provides data strage for one year in case of one measuring value per hour. A CompactFlash® card can be used to allow easy transfer of data to a PC.
The cell is cleaned continuously by the wiper to eliminate any interference in the measurement light path. Therefore, errors in readings from dirty cells, or differences in results before and after cleaning are not evident.
The sample error alarm provided as an option on the previous OPSA120 is now a standard feature.
The mimimum accuracy of analysis ability has been increased to 0.0001 Abs, a ten-fold increase over the previous OPSA-120 unit. Measurements can be made on concentrations as low as 0 to 0.1 Abs.
Product name | Organic Pollutant Monitor (UV meter) |
---|---|
Model | OPSA-150 |
Measurement item | UV absorbance, VIS absorbance, COD conversion concentration, TURB (turbidity) conversion concentration |
Measurement principle | Dual optical path, Dual wavelength, rotary cell length modulation method |
Measurement wavelength | Ultraviolet: 253.7 nm, Visible light: 546.1 nm |
Structure of analyzer unit | Circulation type |
Measurement range (converted by cell 10mm) | UV absorbance/VIS absorbance Full scale: 0.1 Abs to 5.0 Abs (setting by 0.1 Abs is possible) |
Minimum scale | 0.0001 Abs (changeable to 0.001 Abs) |
Repeatability | Within ±2.0% of full scale (Within ±5.0% of full scale for 2.6 Abs to 5.0 Abs) |
Linearity | Within ±2.0% of full scale (Within ±5.0% of full scale for 2.6 Abs to 5.0 Abs) |
Stability | Within ±2.0%/24 h of full scale (Within ±4.0%/24 h of full scale for 2.6 Abs to 5.0 Abs) |
Response time | T90, within 1 minute (flow rate of approximately 5 L/min) |
Cleaning method (cleaning cycle) | Automatic cleaning by wiper (automatic continuous cleaning) |
Display method | LCD display: 320 × 240 monochrome graphics LCD data display with dot backlighting (a touch-panel) Display content: UV absorbance, VIS absorbance, UV-VIS absorbance, COD conversion concentration, TURB (turbidity) conversion concentration |
Calibration method | Solution calibration using zero solution and span solution (calibration solution in the ampoule) (one-touch calibration is possible) |
Sample water condition | Temperature: 2°C to 40°C (should not be frozen) Flow rate of sample water: Min. 2 L/min, Max. 20 L/min * Keep the unit warm (or preserve the heat) to ensure protection against frozen sample water in the cold weather. |
Ambient temperature and humidity | Ambient temperature: 0°C to 40°C, Ambient humidity: 85% or below |
Analog output | Point: 3 Type: arbitrary 3 items from UV absorbance, VIS absorbance, UV-VIS absorbance, COD conversion concentration, and TURB (turbidity) conversion concentration (selectable) Specification: 4 mA to 20 mA DC or 0 mA to 16 mA DC, insulated output (however, non-insulated among each channel) Maximum load resistance: 600 Ω |
Contact output | Point: 6 Type: arbitrary 4 items from POWER OFF, MAINTENANCE (standard, fixed allocation) BATCH ALARM, COD MAX ALARM, TURB MAX ALARM (standard), LIGHT SOURCE MALFUNCTION, NO SAMPLE, CLEANING MOTOR ERROR, ANALYZER ERROR (selectable) Content: POWER OFF: when the power OFF occurred. BATCH ALARM: CLEANING MOTOR ERROR, LIGHT SOURCE MALFUNCTION, ANALYZER ERROR MAINTENANCE: at in-maintenance mode and calibration mode, and when maintenance switch is ON. Specification: contact output without voltage, a contact Contact rating: 125 V AC/0.3 A, 30 V DC/1 A (however, for load resistance) each output COM-independent |
Contact input | Point: 2 Type: Float switch input, time adjustment input Specification: contact input without voltage (open collector can be connected), insulated input ON resistance: Max. 100 Ω, Open voltage: 5.5 V DC, Short circuit current: Max. 10 mA |
Communication | Interface: RS-232C compliant Communication speed: 19200 bps |
Data memory | Memorize the data of measurement value of the measurement item in the main body. Memorized data can be transferred to the CF card. Memory interval: 1 minute or 1 hour Memory time: on the hour every hour Data memory time: interval of 1 minute: approximately 10 days interval of 1 hour: approximately 1 year * The latest data is memorized. |
Light source/detector | Light source: low pressure mercury-vapor lamp, detector: silicon photo diode |
Piping sockets |
|
Structure | Weather-proofed for outdoor installation (equivalent to IP54) |
Main material of each unit | SUS, PVC, PP, CR, SiO2 |
Power | 100 V to 230 V AC ±10%, 50/60 Hz |
Power consumption | 100 V to 120 V AC, Max. 45 VA 200 V to 230 V AC, Max. 60 VA |
Mass | Control unit: approximately 5.0 kg Analyzer unit: approximately 5.6 kg |
Dimension | Control unit: 240 (W) × 104 (D) × 320 (H) mm Analyzer unit: 200 (W) × 180 (D) × 403 (H) mm (Protruding portion are not included.) |
Painting color | Control unit: Munsel 5PB 8/1 |
Conditions for installation |
|
Do you have any questions or requests? Use this form to contact our specialists. * These fields are mandatory.
Automated Organic Analysis & Early Warning Sentinel
For Instant Water Quality Reports
Stand-alone Type Chemical Concentration Monitor
Fiber Optic Type Chemical Solution Concentration Monitor
Carbon Sensor Conductivity Meter (Low concentration type)
KOH Monitor
Wide range TMAH Concentration Monitor
High Precision TMAH Concentration Monitor
Flat Carbon Sensor Conductivity meter
Carbon Sensor Resistivity Meter
2-Channel Resistivity Meter
Field-installation Type Turbidity Meter (Four-Wire Transmission, High-sensitivity)
Field-installation Type Turbidity Meter (Four-Wire Transmission, Immersion Type)
Field-installation Type Turbidity Meter (Four-Wire Transmission, For Wide Range)
H2O2 Monitor
For your specific application
Affordable high performance ICP-OES
Stand-alone Type Chemical Concentration Monitor
Fiber Optic Type Chemical Solution Concentration Monitor
Field-installation Type Turbidity Meter (Four-Wire Transmission, High-sensitivity)
Field-installation Type Turbidity Meter (Four-Wire Transmission, Immersion Type)
Field-installation Type Turbidity Meter (Four-Wire Transmission, For Wide Range)
H2O2 Monitor