CCD Sensor Format | 2048 × 70 |
Quantum Efficiency at 20 °C | See type A curve for VUV and UV-VIS QE responses |
Pixel Size | 14 μm × 14 μm |
Image Area | 28.7 mm × 0.98 mm; 100% fill factor |
Deep Thermoelectric Cooling | -50°C under vacuum (10-6 mbar) with +25°C ambient or -30°C under N2 purging |
Single Pixel Well Capacity | 50 000 e¯/pixel (minimum); 60 000 e¯/pixel (typical) |
Serial Register Full Well Capacity | 250 000 e¯/pixel (minimum) 500 000 e¯/pixel (typical output register saturation) |
Scan Rates | 45 kHz and 500 kHz |
Readout Noise (at 45 kHz and at -50 °C)*1 | 9 e– (typical) to 12 e¯ (maximum) |
Readout Noise (at 500 kHz and at -50 °C)*1 | 20 e- (Typical) to 25 e- (Maximum) |
Maximum Spectral Rate | 20 Hz at 45 kHz scan rate 189 Hz at 500 kHz scan rate |
Digitization | 16 bit ADC |
Dynamic Range (Typical for Single Pixel)*2 | 55 500:1 |
Non Linearity (Measured on Each Camera) | <0.15% (typical) at 45 kHz (0.4% maximum) <0.20% (typical) at 500 kHz (1% maximum) |
Dark Current at -50 °C*3 (Note that pixel size = 14 μm) | 0.05 e – /pixel/s (typical) |
Software-Adjustable Gains | 2,4 and 10 e-/count at –50°C |
Environmental Conditions | - Operating Temperature 0°C to 40°C ambient
- Relative Humidity < 70% (non-condensing)
- Storage Temperature -25°C to 50°C
|
Weight | 1.769 kg (3.90 lb) |
Dimensions | Refer to mechanical drawings |
Power Requirements AC-DC Power Supply (Provided) Recommendation for OEM Supplying Camera Power Directly: | 90- 264 Vac, 47 - 63 Hz - Pin:+9V, ± 5%, 6.44A maximum
- Regulation:+8.55 Vmin, +9Vtyp, +9.45 Vmax
- Ripple & Noise:200 mVpp maximum
|
Minimum Computer Requirements: | - 3.0 GHz single core or 2.4 GHz multi core processor
- 2 GB RAM
- 32 bit or 64 bit compatible
- 500 MB free hard disc space (additional disc space may be required depending on data storage needs)
- USB 2.0 High Speed Host Controller capable of sustained rate of 40 MB/s
- Windows (XP, Vista and 7)
|