Measurement of Sulfur and Chlorine in Petroleum Products
The MESA-7220 is based on the latest advances in Energy Dispersive X-ray Fluorescence (EDXRF) technology. It utilizes a proprietary X-ray optical technology and produces a polarized, monochromatic X-ray source. This approach is critical for a user to achieve an ultra low noise background for the best limits of detection for sulfur and chlorine. This provides excellent performance at high concentration (multiple wt.% levels), simultaneous multi-element analysis and applicability over a wide range of sample types, and excellent low ppm capabilities.
*Available in the Americas. For additional information, please contact us.
• Simultaneous elemental analysis.
• One calibration – for multiple petroleum matrices over a wide dynamic range.
• Dynamic Analysis Range: Sulfur: 0.7 mg/kg – 10.0 wt% and Chlorine: 0.6 mg/kg – 10.0 wt%
• Various sample types including liquids, solids,pastes, powders, films, and pellets.
• No purge or combustion gases required.
• Correction for oxygen interference that can effect sulfur results significantly in other analyzers.
• User defined measurement times (90 - 600 seconds.)
• Single digit ppm repeatability achieved on low level samples.
• Multiple Calibration Curve capability with easy storage and retrieval for matrix matching.
• Rugged, user serviceable Kapton Windows.
• Multiple external outputs: Printer, serial I/O, USB (2) and network (RJ-45), (2 USB ports, side and back. PS2 connects, ports for mouse and keyboard,ethernet connections.)
• Accurately measures sulfur and chlorine in the presence of moisture.
X-ray Tube | Pelletier-cooled, Ag anode X-ray tube; 0-12.5kV; 0-2mA |
Polarizing Mono-chromater | Closed-coupled, doubly curved HOPG crystal reflects Ag Ka radiation for optimal excitation of S and Cl. |
Si drift Detector | Stable peak positions and spectral resolution (<150 eV) up to an input count rate of 200 kcps. |
Dynamic Range | 1.0 mg/kg - 10% (wt/wt) - (1 ppm - 100,000 ppm). The sample chamber is under vacuum to enhance sensitivity. |
Vacuum Chamber Window | User replaceable 7.5 mm Kapton window. The detector is also protected by a replaceable safety window between it and the sample cup window. |
Sample volume | 5 mL |
Included Accessories and Options | • Printer • Keyboard and mouse |
Software | • Create unlimited number of calibrations. • On screen diagnostics of all operating parameters. • An intuitive touch screen based, user interface. • Ability to easily add new functionality. |
Operating Requirements | • Nominal input voltage: 120-240 V +/- 10%AC, 50/60Hz, 125W |
Computer | • Simple, touch screen interface with a sealed, field-hardened 8.4” screen (800 x 600 px) |
Weight | 21kg (45 lbs.) |
Dimensions | 330 mm (Width) x 508 mm (Depth) x 413 mm (Height) (13 x 20 x 16.25 inches) |
Do you have any questions or requests? Use this form to contact our specialists. * These fields are mandatory.
X-ray Fluorescence Sulfur/Chlorine-in-oil Analyzer
X-ray Fluorescence Sulfur-in-Oil Analyzers
Total Nitrogen and Sulfur Analyzer
Total Halogen and Sulfur Analyzer
X-ray Fluorescence Sulfur/Chlorine-in-oil Analyzer
X-ray Fluorescence Sulfur-in-Oil Analyzers
X-ray Analytical Microscope Super Large Chamber Model