Tocadero EVO

Fast, reagent-free, continuous online TOC analyzer

The Tocadero EVO TOC meter is your precise solution for monitoring organic contamination in ultrapure and pure water—ideal for the pharmaceutical industry, microelectronics, power plants, and chemical production. With innovative UV oxidation and differential conductivity measurement, the device detects even the slightest organic impurities—in real time.

Typical areas of application for the TOC meter:

  • Ultrapure water monitoring in semiconductor production
  • Pharmaceutical water quality control (PW, HPW, WFI)
  • Monitoring of rinse water in CIP applications
  • Laboratory applications with single sample analysis
  • Water quality control for electrolysers

 

Fast online measurement process with response times in the seconds range, an offline mode with optional autosampler, and reagent-free analysis for maximum cost-effectiveness and minimal operating costs. The integrated software complies with 21 CFR Part 11 and supports you with audit trails, logging, and regulatory compliance.

Use of the TOC meter in a regulated environment

The Tocadero EVO TOC meter is specially designed for use in regulated industries such as the pharmaceutical industry. It helps you comply with international requirements, including 21 CFR Part 11, with a compliant software solution featuring audit trail, log, and tracking functions. This ensures traceable data integrity and compliance with the requirements of common pharmacopoeias such as USP and Ph. Eur. Thanks to its reliable differential conductivity measurement, the device is ideal for monitoring PW, HPW, and WFI in GMP-regulated processes.

What makes this TOC meter unique

  • Results in seconds: Continuous online measurement for immediate TOC values.
  • Reagent-free: No chemicals required – safe, sustainable, and cost-effective.
  • 21 CFR Part 11 ready: Ideal for regulated industries such as pharmaceuticals – including audit trail and data security.
  • Dual mode: Online monitoring plus offline measurements with optional autosampler.
  • Long-lasting UV technology: Maximum measurement accuracy with minimal maintenance.
  • Automatic dilution: Lower consumption, precise calibration, easy operation.
  • Compact & low maintenance: Saves space, time, and resources during operation.

 

Easy calibration with integrated sample pump and dilution

No additional equipment is required to calibrate the Tocadero EVO. The integrated sample pump automatically feeds the calibration standard into the device. The different concentration levels are produced and measured autonomously using the dilution function. This saves time and money on additional equipment or standards and gives complete flexibility in terms of when and how the Tocadero EVO is calibrated.

High comparability of laboratory and online measurements

Individual samples often have to be taken during the course of the process and measured in the laboratory. The comparability between laboratory and online measurements is often questionable. However, if the same system is used for online and laboratory measurements, comparability is guaranteed 100%.

Applications

IndustryApplication example
Pharmaceuticals, medical technology & biotechnologyTOC monitoring in WFI production systems, TOC control in cleaning validation, TOC measurement in cooling water circuits, and TOC monitoring in pharmaceutical wastewater.
Semiconductor & Electronics manufacturingTOC monitoring in ultrapure water systems, TOC control in process chemicals, TOC measurement in CMP slurries, and TOC monitoring in semiconductor wastewater.
Seawater desalinationTOC measurement in seawater inflow, TOC control in membrane cleaning solutions, and TOC monitoring in concentrate effluent.
Hydrogen productionTOC control in electrolysis water, TOC monitoring in cooling circuits, TOC measurement in electrolyte solutions, and TOC monitoring in process wastewater.
Energy generation/ Power plantsTOC monitoring in boiler feed water, TOC control in cooling water systems, TOC measurement in condensates, and TOC monitoring in power plant wastewater.
Drinking water treatmentTOC measurement in raw water, TOC monitoring after treatment stages, TOC control in pure water, and TOC monitoring in municipal wastewater.
Cosmetics & Personal Care IndustryTOC control in production water, TOC monitoring in cooling systems, TOC measurement in process solutions, and TOC monitoring in cosmetic wastewater.
Chemical industryTOC monitoring in reactor cooling water, TOC control in process water, TOC measurement in catalyst suspensions, and TOC monitoring in chemical wastewater.

Service and Support

When service is required, you will be assisted by a qualified contact person – not a call center. If necessary, a technician will be on site within 48 hours. All maintenance information is provided clearly, comprehensively, and directly by the system – for maximum transparency and minimal downtime.

Made in Germany – Economical, Sustainable, Durable

The Tocadero EVO is manufactured in Berlin. For you, this means short delivery routes, fair prices, and long-term supply security. Consumables and replacement materials are inexpensive, and maintenance costs are low—a well-designed solution for cost-conscious operators.

Specifications

Methods & DetectionContinuous flow with UV oxidation and conductivity detection
Cycle Time1 sec. (online method)
1 min. + 1 min. rinsing (offline method)
Sample Flowapprox. 14 ml/min
Measurement Range0 - 1000 ppb TOC
Application-dependent up to 2000 ppb TOC
Accuracy & PrecisionLimit of detection: 0,1 ppb
Accuracy: ± 0,05 ppb online
Connections1 x sample, 1 x check
1 x sample, 3 x check (optional)
NormalizationsUSP <643>, EP 2.2.44, JP, CP
FDA 21 CFR Part 11 - Compliance for Pharma
Environmental ConditionsRelative humidity: < 80 % non condensing
Temperature: 10 - 40 °C
Sample TypePure Water (PW) & Ultra Pure Water (UPW),
Water for Injection (WFI), Boiler Feed Water
Conductivity Sample2 μS/cm
Application-dependent up to < 10 μS/cm
Input Pressure< 0,5 bar
Sample Temperature< 50 °C
Interfaces2 x analog (0/4 - 20 mA), according NAMUR NE43
1 x relais (collective error)
RJ45 Ethernet, USB
Optional: 2 additional analog outputs, 2 additional relay outputs, autosampler
Dimensions338 W x 200 D x 450 H mm
Weight11 kg
Power110 - 230 VAC ± 10 %, 47 - 63 Hz, 60 Watt

 

Informationsanfrage

Sie haben Fragen oder Wünsche? Nutzen Sie dieses Formular, um mit unseren Spezialisten in Kontakt zu treten.

* Diese Felder sind Pflichtfelder.

Verwandte Produkte

HT-110
HT-110

On-line TOC analyzer

TOCADERO ONE
TOCADERO ONE

Automatic Total Organic Carbon Analyzer (On-line TOC analyzer)

Corporate