Model | ENDA-5000 | ||||||
Component | NOX | SO2 | CO | CO2 | O2*1 | ||
Measurement methods | NDIR | NDIR | NDIR | NDIR | Magneto-pneumatic detection | ||
Range*2 | Standard | 200~5000 ppm | 200~5000 ppm | 200~5000 ppm | 5~25 vol% | 10~25 vol% | |
| Optional | 100 ppm~ | 50 ppm~ | 100 ppm~ | - | - | |
Range Ratio | Within a factor of 10 | Within a factor of 10 | Within a factor of 10 | Within a factor of 5 | Within a factor of 2.5 | ||
Repeatability | Within 0.5% of full scale (with optional range, or during O2 measurement, ±1.0% of full scale) | ||||||
Linearity (indicator error) | ±1.0% of full scale | ||||||
Zero drift | ±1.0% of full scale (assuming surrounding temperature is maintained within 5 °C) | ||||||
Span drift | ±2.0% of full scale/week (assuming surrounding temperature is maintained within 5°) | ||||||
Response time*3 | Within 60 seconds (Td + T90 from equipment intake area) (sample flow 0.6 L/min.) (within 240 seconds for SO2 only) | ||||||
Overall interference frrom co-existing gases | ±2.0% of full scale/week (within standard range, with standard gas formation) | ||||||
Display | Touch panel LCD (backlight) (four usable lines) | ||||||
Environment | Temperature | -5 to 40 °C (away from direct sunlight and radiation heat )*4 | |||||
Humidity | 85% or less (no condensation) | ||||||
Vibration | 100 Hz, 0.3 m/S2 or less | ||||||
Dust | Standard environment or better | ||||||
Measuring Gas Condition | Temperature | 250 °C or lower | |||||
Dust | 0.1 g/Nm3 or less | ||||||
Standard gas composition*5 | NO: 500 ppm or less; NO2: 15 ppm or less; SO2: 1000 ppm or less; SO3: 50 ppm or less; | ||||||
Sampling method | Dry sampling using an electric cooler | ||||||
Sample gas flow | 2.5 L/min~3.0 L/min | ||||||
Sample inlet tube | Teflon tubing (Ø8/Ø6 mm) | ||||||
Sample gas pressure | ±4.9 kPa (three points selected) |
| |||||
Pressure control | Pressure control uses a regulator and cylinder; Reduced pressure sampling; Control pressure: -4.9 kPa | ||||||
Output | DC 4 to 20 mA (absolute output) (DC 0 to 16 mA/DC 0 to 1 V/DC 1 to 5V optional) Max. 12 output systems | ||||||
External output | Analysis alerts, analysis warnings, range display, corrections, conservation, purging (option) | ||||||
Correction method | Dry correction, automatic correction (correction cycle: 7 days standard, can be adjusted to between 1 and 99 days), manual correction | ||||||
Calibration gas | Zero gas | With measurement method authorization: N2, When there is no measurement method authorization: N2 or ambient air | |||||
O2 carrier gas | Ambient air | ||||||
Span gas | Gas cylinder for each component measured (when there is no measurement method authorization: O2 or ambient air can be used) | ||||||
Probe collection point filter | Flange: JIS 10K, 40 AFF; Sample probe tube length: 1000 mm; Material: SUS-316 stainless steel; | ||||||
Power supply | AC 100 V ±15 V(85 V~115 V) | ||||||
Power frequency | 50/60 Hz (switchable) | ||||||
Power consumption | About 800 VA (heated piping 30m: +1100 VA; heater in tray: +300 VA) | ||||||
Exterior dimensions/Mass | 600 (W) x 1770 (H) x 300 (D) mm (high pressure gas cylinders, 3.4 L cylinders, maximum of 3 cylinders); About 180 kg (not including cylinders) | ||||||
Materials in contract with sample gas | SUS-316 stainless steel, SUS-304 stainless steel, Teflon, polypropylene, polyethylene, | ||||||
Enclosure | Independent outdoor installation | ||||||
Color/Finish | Semi-gloss Munsell 5Y7/1 on all inner and outer surfaces |