Strain Measurements of a Si Cap Layer Deposited on a SiGe Substrate, Determination of Ge Content

Typical structure of Si/SiGe/Si sample.

Mechanical strain directly affects the frequency positions of the Raman modes and can lift their degeneracy. Typically, it is possible to observe and detect strain by analyzing the shift in the band position, but it can also affect the band shape and induce broadening and deformation peaks.

Raman spectroscopy is one of the most popular tools for investigating the basic properties of semiconductors. It is particularly efficient in establishing the characteristics of microelectronic devices because the performance of SiGe based devices highly depends on the knowledge of the composition, Ge content and the strain of the layers. Indeed, the ability to measure alloy composition and strain in semiconductor structures is essential for the calibration of growth processes and control of the electrical and optical behavior of these materials.

The first requirement for the study of mechanical stress with Raman spectroscopy is that the investigated material exhibits Raman active modes, ie that there is a well defined Raman band in the spectrum. These include materials such as diamonds, silicon, SiGe, InGaAs, GaAs, GaN, to name a few. Mechanical strain directly affects the frequency positions of the Raman modes and can lift their degeneracy.

Typically, it is possible to observe and detect strain by analyzing the shift in the band position, but it can also affect the band shape and induce broadening and deformation peaks. These effects depend upon the material characteristics and on the stress and strain geometry. In this study, we focus on the choice of the laser excitation and measurement conditions. Results obtained in the visible and in the UV range are presented.

Application downloads:

Related products

LabRAM Odyssey
LabRAM Odyssey

Confocal Raman & High-Resolution Spectrometer

제품 문의

HORIBA제품의 자세한 정보를 원하시면, 아래의 양식에 내용을 입력을 부탁드립니다.

* 는 필수입력항목입니다.

Corporate