Name | Tunable Laser Gas Analyzer |
Model | TX-100 |
Measurment target | HCl and H2O in process gas |
Measurement Component | HCl, or HCl and H2O simultaneous measurement |
Measuement principle | NDIR (Tunable Laser Absorption Spectroscopy method) |
Measurement type | Probe type |
Laser class | CLASS 1 (IEC60825-1:2014) |
Protection degree | IP65 |
Materials | SUS316, SUS304, BK7, FKM, PTFE, Aluminum alloy |
Tube connections | Rc1/8 (Tube of 8 mm O.D. /6 mm I.D.) |
External dimensions | Analyzer unit: 358(W) x 517(D) x 453(H) mm |
Mass | Analyzer unit: 19 kg, Probe: 20 kg |
Flange specifications | JIS10K100AFF, JIS5K125AFF, ANSI150Lb4BRF |
Power supply voltage | 24 V DC±10% |
Power consumption | Max. 100W |
Display | LCD (128 x 64 dots), LED (3 units: Status display) |
Display contents | Gas concentration, ransmission, prcess gas temperature, Process gas pressure, |
Communication | ModbusTM/TCP |
Analog input | Two channels (Process gas temperature, process gas pressure) 4-20 mA (isolated input) |
Analog output | Two channels (Process gas concentration*, transmission, process gas temperature, process gas pressure) 4 -20 mA (Isolated output) Max. load: 550Ω |
Digital input | One channel (External alarm) Non-voltage digital input (Isolated input) Input open signal: 100 kΩ or more Input close signal: 200 Ω or less |
Digital output | Two channels (Warning/Calibration/Warm-up/Maintenance state/Fault) |
*고정값 또는 실시간 HCl 보정을 H2O 농도로 선택 가능
-Modbus는 슈나이더 전기(Schneider Electric Inc.)의 트레이드 마크입니다.
-특수한 프로브 길이가 필요한 경우 호리바에 문의 바람
Performance
Measurement component | HCl | H2O |
Measurement range (250℃、101.3kPa) | 0-50,100,200,300,400,500 ppm*1 | 0-50 vol% |
Measured optical path length | 2m | |
Repeatability | Within±1.0 % of full scale | Within ±2.0 % of full scale |
Linearity | Within±1.0 % of full scale | Within ±2.0 % of full scale |
Zero drift | ±2.0% of full scale / 6 months | |
Span drift | ±2.0% of full scale / 6 months | |
Data update cycle | Approx. 2sec | |
Recommended | 6 months | |
Calibration method | HCl: Manual calibration using gas cell H2O: Factory calibration*2 | |
Interference | Within±2.0% of full scale | |
Warm-up time | 3hours |
*1: Select a measurement range.
*2: You can adjust the H2O indicated value to the manual analysis value. Consult HORIBA for more details.
Usage Conditions
Ambient temperature | -20 ~ 55°C (no direct sun rays, radiant heat and heat conduction from flange) |
Ambient humidity | Relative humidity: 90% or less (no condensation) |
Dust | Ambient criterion or less |
Air purge | Instrument air (must not contain oil or mist; no dew formation) |
Air purge flow rate | 60 L/min or more (depend on the process gas conditions) |
Vibration | At 1 to 200 Hz, 1.96 m/s2 or less (excluding resonant frequency) |
Process gas conditions | Temperature: 150 to 250°C Pressure: 101.3 ± 10 kPa Flow speed: 25m/s Dust: 0.1mg/Nm3 or less |
외형도
측정 기술
프로브 유형의 광학 시스템 작동 방법
분석기의 동일한 위치에서 광원과 검출기는 스택에 삽입되는 프로브에 부착되며 프로브 끝에 설치된 거울은 프로브를 통해 레이저 광선을 수신하여 검출기로 다시 반사합니다.
<측정>
영점 점검 및 측정 교대 수행
<교정>
교정 셀에 스팬 가스를 유입한 후 교정 수행
프로브를 스택에 부착한 상태로 보정 완료
TX-100의 경우 데이터 관리는 분석 단위 디스플레이, HMI(Human Machine Interfce) 유닛 또는 PC-HMI 소프트웨어를 포함한 유형 중에서 선택 가능합니다.
악세사리
PC-HMI 소프트웨어
PC-HMI 소프트웨어는 윈도 PC로 조작할 수 있는 무료 소프트웨어로 이더넷을 통해 분석기가 접근할 수 있습니다. 화면에는 농도, 투과율, 트렌드 그래프와 같은 여러 가지 지표와 교정을 포함한 다양한 작동 및 설정이 표시됩니다.
PC 운영 환경
PC-HMI 소프트웨어를 PC에 설치하기 전에 운영 체제가 아래 조건을 충족시키는지 확인하세요. Windows 7(32/64 비트)/8.1(32/64 비트)/10(32/64 비트)과 호환되는 PC/AT.
다운로드 방법
아래의 'TX-100 PC-HMI 소프트웨어 다운로드' 를 클릭하시면 나오는 설문지 페이지에 내용을 기입하고 보내기 버튼을 클릭하면 소프트웨어 다운로드 URL이 이메일 주소로 전송됩니다.
* 해당 무료 소프트웨어는 프로모션용으로만 제공되며 호리바는 소프트웨어의 다운로드 또는 사용으로 인한 손상 또는 손실에 대해 책임을 지지 않습니다.
Download TX-100 PC-HMI Software
트레이드마크
윈도(Windows)는 미국, 일본 등지에서 마이크로소프트사의 상표 혹은 등록 상표이며 이더넷은 Fuji Xerox Co., Ltd. 상표입니다.