HC-200N

Ammonia/Nitrate Nitrogen Meter

Ammonia nitrogen and nitrate nitrogen concentrations in the tank are measured in real time with an ion electrode method sensor.

Reduces power consumption and costs at sewage treatment plants and other water treatment facilities. The proprietary technology enables high durability and long-term stability.

 

NH4-NNO3-N

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

Simultaneous measurement of two components

Enables simultaneous measurement of ammonia nitrogen and nitrate nitrogen concentrations, leading to reduced power consumption and costs at water treatment facilities such as nitrification-denitrification tanks.

Long-term stability

The sensor's unique protective structure and optimized internal solution enable stable measurements. 

Easy maintenance

Replaceable chip without the need for tools. No need to call manufacturer maintenance.

Prevention of contamination for increased longevity

HORIBA's proprietary protective film provides sensor durability. Combined with an ultrasonic cleaner, antifoulding properties are improved.

Reduced risk of sudden operational shutdowns

The self-diagnostic function of the sensor enables the immediate detection of anomalies.
*Ammonia nitrogen only.

 

HC-200N Specifications

Product nameAmmonia/Nitrate Nitrogen Meter
Converter modelHC-200N

 

Ammonia nitrogen

Sensor modelAM-2000-A: Ammonia nitrogen sensor
Sensor tip model7691: Ammonium ion tip, 7692: Potassium ion tip for compensation, 7211: Reference tip
Measurement rangeConcentration: 0 mg/L to 1000 mg/L (display range: 0 mg/L to 2000 mg/L)
Temperature: 0℃ to 40℃ (display range: -10℃ to 110℃)
Display resolutionConcentration: 0.01 mg/L: 0.00 mg/L to 10.00 mg/, 0.1 mg/L: 0.0 mg/L to 100.0 mg/L, 1 mg/L: 0 mg/L to 1000mg/L
Temperature: 0.1℃
RepeatabilityConcentration: ±3% or ±0.2 mg/L, whichever is greater for the readings (standard solution)
Temperature: ±0.3℃(equivalent input)
Temperature 
compensation
Compatible temperature element: platinum resistor: 1 kΩ (0℃)
Temperature measurement range: 0℃ to 40℃
Temperature calibration function: 1 point calibration for comparison with the reference thermometer
Potassium ion
concentration
compensation
Compensation range: the potassium ion concentration is not more than 10 times the ammonium ion concentration and 1000 ppm or less.
Compensation error: ±20%(reading)
Measurable pH rangepH4 to pH8
CalibrationReference solution calibration of the ammonia nitrogen (2points)
Type of the reference solution: select 1 mg/L to 10 mg/L or 10 mg/L to 100 mg/L.
Additional functions: automatic defect calibration judgment, calibration history
AdjustmentSample adjustment using hand analysis (1 point, 2 points*1), standard curve input function (primary system)
Additional functions: adjustment history (number of days elapsed from the previous adjustment, sample adjustment coefficient)
Abnormal detectionCalibration error, sensor diagnosis error, converter error

*1 When AM-2000 is connected, the 2-point sample adjustment function becomes unavailable.

 

Nitrate nitrogen

Sensor modelNO-2000: Nitrate nitrogen sensor
Sensor tip model7693: Nitrate ion tip, 7694: Chloride ion tip for compensation, 7211: Reference tip
Measurement rangeConcentration: 0 mg/L to 1000 mg/L (display range: 0 mg/L to 2000 mg/L)
Temperature: 0℃ to 40℃ (display range: -10℃ to 110℃)
Display resolutionConcentration: 0.01 mg/L: 0.00 mg/L to 10.00 mg/, 0.1 mg/L: 0.0 mg/L to 100.0 mg/L, 1 mg/L: 0 mg/L to 1000mg/L
Temperature: 0.1℃
RepeatabilityConcentration: ±3% or ±0.2 mg/L, whichever is greater for the readings (standard solution at 25℃)
Temperature: ±0.3℃(equivalent input)
Temperature 
compensation
Compatible temperature element: platinum resistor: 1 kΩ (0℃)
Temperature measurement range: 0℃ to 40℃
Temperature calibration function: 1 point calibration for comparison with the reference thermometer
Chloride ion
concentration
compensation
Compensation range: when the chloride ion concentration is 100 times greater than the nitrate ion concentration, a correction factor must be applied to account for interference.
Compensation error: ±20%(reading)
Measurable pH rangepH2 to pH12
When the pH is between pH2 to pH4 or pH9 to pH12, use sensor cap AM-CAP-PVC.
Chloride tip can be used in the pH2 to pH8. However, when used at pH2 to pH4, the lifetime of the tip may be shortened.
Measurement inhibitorsComponent: if the surfactant concentration is 10mg/L or higher, sensor performance may degrade, and measurement may not be possible. 
CalibrationReference solution calibration of the ammonia nitrogen (2points)
Type of the reference solution: select 1 mg/L to 10 mg/L or 10 mg/L to 100 mg/L.
Additional functions: automatic defect calibration judgment, Calibration history 
AdjustmentSample adjustment using hand analysis (1 point, 2 point), standard curve input function (primary system)
Additional functions: adjustment history (number of days elapsed from the previous adjustment, sample adjustment coefficient)
Abnormal detectionCalibration error, sensor diagnosis error, converter error

 

Common specifications (Converter)

Transmission output3 points, 4 mA to 20 mA DC, input/output isolation type, 900Ω(max)
Select 3 items from below.
Ammonia nitrogen concentration: arbitrary setting within the measurement range is possible.
Nitrate nitrogen concentration: arbitrary setting within the measurement range is possible.
Chloride concentration: arbitrary setting within the measurement range is possible.
Temperature reading of the ammonia nitrogen meter: arbitrary setting within the range from 
⁻10℃ to 110℃ is possible.
Temperature reading of the nitrate nitrogen meter: arbitrary setting within the range from 
⁻10℃ to 110℃ is possible.
Contact output5 points, No-voltage contact output, relay contact, SPST(1a),SPDT(1c)
 R1,R2: relay contact SPST(1a)
           Contact function: selectable from among the upper limit alarm, lower limit alarm, during
           transmission output hold, and cleaning output (closed at alarm operation, open under 
           normal conditions and in a power off state).
 FAIL: relay contact SPDT(1c)
           Contact function: error alarms for values outside the measurement range and self 
           check output can be set. 
           Delay time: 0 seconds to 600 seconds
 R3,R4: relay contact SPST(1a)
Cleaning output2 points, Voltage contact output (connection power supply voltage output)
Relay contact, SPDT(1c), SPST(1a)
Contact input1 point, Contact type: open collector no-voltage “a” contact
Operating temperature range⁻20℃ to 55℃(no freezing)
Operating humidity rangeRelative humidity: 5% to 90%(no condensation)
Storage temperature⁻25℃ to 65℃ 
Power supply100V to 240V AC ±10% 50/60Hz, Power consumption: 28 VA(max)
Compatible standardsCE marking, FCC Rules
StructureOutdoor installation type, IP65, 50 A pole-mounted or wall-mounted, material of case: aluminium alloy, material of the mounting brackets and hood :SUS304
External dimensions180(W)×155(H)×115(D)mm (excluding mounting brackets)
MassMainframe: approx. 3.5kg, hood, mounting bracket: approx. 1kg

 

Ammonia nitrogen sensor

ModelAM-2000-A
Measurement rangeTemperature: 0℃ to 40℃ (no freezing)
Measurement depth0m to 10m
MaterialSUS316, FKM, PVC
MassApprox. 2.7kg (including 10m cable)

 

Nitrate nitrogen sensor

ModelNO-2000
Measurement rangeTemperature: 0℃ to 40℃ (no freezing)
Measurement depth0m to 10m
MaterialSUS316, FKM, PVC, POM
MassApprox. 2.7kg (including 10m cable)