APSA-370/CU-1

Ambient Hydrogen Sulfide Monitor

APSA-370/CU-1 Ambient H2S monitor measures SO2 converted H2S through oxidation catalyst based on Ultra Violet Fluorescence. In order to reduce measurement error by water concentration fluctuatin in ambient, humidifier is set at sample line. It enables long-term stability according to stabilizing catalytic reaction.

Segment: Process And Environment
Division: Ambient
Base product
Manufacturing Company: HORIBA, Ltd.
  • Monitor H2S with high sensitivity range in ambient air continuously.
  • Set humidifier at sample line to satbilize catalytic reaction.

 

 

Measurement component 

Hydrogen sulfide (H2S) in ambient air
Measurement principleOxidation catalyst + Ultra Violet Fluorescence
Measurement range0-0.1/0.2/0.5/1.0 ppm
Lower detectable limit2 ppb
Repeatability±3.0% of F.S.
Linearity±2.0% of F.S.
Zero drift±2% of F.S./day
Span drift±5% of F.S./day
Response time (T90)180 sec. or less (from CU-1converter unit)
Sample gas flow rateapprox. 0.7L/min
Operation gasNone
IndicationMeasuring value, alarm
AlarmAPSA-370: Calibration error, Battery error, Flow rate error, Pressure error, Temperature error in catalyzer, etc.
CU-1: Temperature error in catalyzer
Inpt/Output0-1V/0-10V/4-20mA (2 systems: either (1) momentary value and integrated of (2) moving average value)
Contact input/output
RS-232C (option)
Temperature/Humidity0-40°C / 85% or less
Power100/110/115/120/220/230/240 VAC, 50/60Hz
Power consumptionAPSA-370: 150VA
CU-1: 150VA
DimensionAPSA-370: 430(W) x 550(D) x 221(H) mm
CU-1: 430(W) x 550(D) x 221(H) mm
MassAPSA-370: approx. 19kg
CU-1: approx. 10kg

REQUEST FOR INFORMATION

Do you have any questions or requests? Use this form to contact our specialists.

Related products

51 series
More51 series
Explosion-proof Gas Analyzer
AP-370 series
MoreAP-370 series
Air Pollution Monitor
APHA-370
MoreAPHA-370
Ambient THC monitor
APNA-370
MoreAPNA-370
Ambient NOx Monitor
APNA-370/CU-2
MoreAPNA-370/CU-2
Ambient Ammonium Monitor
APOA-370
MoreAPOA-370
Ambient Ozone Monitor
APSA-370
MoreAPSA-370
Ambient Sulfur Dioxide Monitor
AQMS
MoreAQMS
Air Quality Monitoring System
CEMS
MoreCEMS
Continuous Emission Monitoring System
ENDA-5000 series
MoreENDA-5000 series
Stack Gas Analysis System
FTX-ONE
MoreFTX-ONE
FTIR Exhaust Gas Analyzer
GA-370
MoreGA-370
Trace Gas Monitor
GI-700 series
MoreGI-700 series
Stack Gas Analyzer
IR-200
MoreIR-200
Gas Monitor for Chamber Cleaning End Point Monitoring
IR-300
MoreIR-300
Vapor Concentration Monitor
IR-400
MoreIR-400
High-grade type Gas Monitor for Chamber Cleaning End Point Monitoring
MEXA-1170HCLD
MoreMEXA-1170HCLD
Heated Type NOx Analyzer
MEXA-1170HFID
MoreMEXA-1170HFID
Heated Type THC Analyzer
MEXA-1170SX
MoreMEXA-1170SX
Motor Exhaust Gas Sulfur Analyzer
MEXA-584L
MoreMEXA-584L
Automotive Emission Analyzer
MEXA-ONE
MoreMEXA-ONE
Motor Exhaust Gas Analyzer
MEXA-ONE-FT
MoreMEXA-ONE-FT
FTIR Motor Exhaust Gas Analyzer
MEXA-ONE-QL-NX
MoreMEXA-ONE-QL-NX
Laser Spectroscopic Motor Exhaust Gas Analyzer
OBS-ONE GS Unit
MoreOBS-ONE GS Unit
On-board Emissions Measurement System
PG-300 series
MorePG-300 series
Portable Gas Analyzer
Process
MoreProcess
Process Analyzer System
PX-375
MorePX-375
Continuous Particulate Monitor with X-ray Fluorescence
TX-100
MoreTX-100
Tunable Laser Gas Analyzer
VA-5000 / VA-5000WM Series
MoreVA-5000 / VA-5000WM Series
Multi-Component Gas Analyzer