Industrial simplified fluoride ion concentration meter HC-300F (Two-Wire Transmitter)

HC-300F (Two-Wire Transmitter)

 

 

Division: Process Water
Manufacturing Company: HORIBA Advanced Techno Co., Ltd.

 

  • Suitable for fluoride ion monitoring in effluent from factories, e.g., those of semiconductor FPD and glass manufacturers
  • 0 to 10,000 mg/L wide measuring range (10-range switching)

 

 

ModelHC-300F
Combined sensorFluoride ion electrode
Measurement rangeConcentration:
 0 mg/L to 10000 mg/L (Display range: 0 mg/L to 11000 mg/L)
 The selectable range: 10000 ,5000 ,2000 ,1000 ,500 ,200 ,100 ,50 ,20.0 ,10.0 mg/L
Temperature:
 0°C to 100°C (Display range: -10°C to 110°C )
Transmission outputOutput type:
 4 mA to 20 mA DC: input/output isolated type
Load resistance:
 Maximum: 600 Ω Case of 24 V DC power supply
Linearity:
 Within ±0.08 mA (output only)
Contact inputNumber of inputs: 1
Contact type: No-voltage "a" contact for open collector
Temperature compensationApplicable temperature element:
 Platinum resistor: 1 kΩ (0°C)
 Positive relation resistor with temperature: 10kΩ(25°C)
Temperature compensation range: 0 °C to 100°C
CalibrationNumber of calibration points: Selectable from 1, and 2 points
Self-checkCalibration error
Electrode diagnostic error
Meter error
Operating temperature range-20°C to 60°C (without freeze)
Power supplyPower supply voltage range: 24 V DC (21V to 32V)
Power consumption: 0.6 W(max)
Compatible standardsCE marking, FCC Rules
StructureInstallation: Outdoor installation type
Installation method: Mounted on 50A pole or wall
Protection code: IP65 IEC60529, JIS C0920
Case material: Aluminum alloy (coated with epoxy-denatured melamine resin)
External dimensions180 (W) x 155 (H) x 115 (D) mm (excluding the fittings)
MassMain body: Approx. 2.8 kg; cover and fittings: Approx. 1 kg

Request for Information

Do you have any questions or requests? Use this form to contact our specialists. * These fields are mandatory.

Related products

Aqualog
More Aqualog

Water Treatment Plant Analyzer

CODA-500
More CODA-500

Automatic COD Monitor

Aqualog
More Aqualog

Water Treatment Plant Analyzer

Aqualog Datastream
More Aqualog Datastream

For Instant Water Quality Reports

CS-600F
More CS-600F

Fiber Optic Type Chemical Concentration Monitor

FES-310L
More FES-310L

Sanitary Conductivity Sensor (Flow Through type)

FS-01FC-SL Series
More FS-01FC-SL Series

Sanitary Conductivity Sensor (Flow Through type) for Low Concentrations

Gratings - OEM Diffraction Grating
More Gratings - OEM Diffraction Grating

OEM gratings design and production capabilities

HC-200NH
More HC-200NH

Ammonia Nitrogen Meter

HD-200
More HD-200

Industrial dissolved oxygen analyzer (Four-Wire Analyzer)

HD-200FL
More HD-200FL

Industrial dissolved oxygen analyzer (Four-Wire Analyzer)

HD-300
More HD-300

Industrial dissolved oxygen analyzer (Two-Wire Transmitter)

HD-480
More HD-480

Industrial dissolved oxygen analyzer (The Slim48 series)

HE-200C
More HE-200C

Industrial electric conductivity meter (Four-Wire Analyzer)

HE-200R
More HE-200R

Industrial electric resistivity meter (Four-Wire Analyzer)

HE-300C
More HE-300C

Industrial electric conductivity meter (Two-Wire Transmitter)

HE-300R
More HE-300R

Industrial electric resistivity meter (Two-Wire Transmitter)

HE-480C
More HE-480C

Industrial conductivity meter (low concentration type)

HE-480C-GC
More HE-480C-GC

Carbon Sensor Conductivity Meter (Low concentration type)

HE-480H
More HE-480H

Industrial conductivity meter (high concentration type)

HE-480R(W)
More HE-480R(W)

Industrial resistivity meter

HE-960CA
More HE-960CA

Citric Acid Monitor

HE-960CW
More HE-960CW

2-channel Sanitary Conductivity Meter (Low Concentration type)

HE-960H-TM
More HE-960H-TM

Wide range TMAH Concentration Monitor

HE-960H-TM-S
More HE-960H-TM-S

High Precision TMAH Concentration Monitor

HE-960HC
More HE-960HC

Carbon Sensor Conductivity Meter (High concentration type)

HE-960HI
More HE-960HI

Sanitary Conductivity Meter (Wide-range type)

HE-960LC
More HE-960LC

Carbon Sensor Conductivity Meter

HE-960R-GC(W)
More HE-960R-GC(W)

Carbon Sensor Resistivity Meter

HE-960RW
More HE-960RW

Industrial resistivity meter

HE-960RW-GC(W)
More HE-960RW-GC(W)

2-Channel Resistivity Meter

HE-960TM
More HE-960TM

TMAH Conductivity Meter. Perfect for Controlling TMAH Developer Concentrations

HF HD-960L
More HF HD-960L

Dissolved Oxygen Monitor in Low Concentration

HF-960H
More HF-960H

HF / HCl Concentration Monitor

HF-960M
More HF-960M

Low Concentration Type HF/HCl/NH3 Concentration Monitor

HO-200
More HO-200

Industrial ORP meter (Four-Wire Analyzer)

HP-200
More HP-200

Industrial pH meter (Four-Wire Analyzer)