| Model | HF-960H |
|---|---|
| Measurement method | 4-electrode carbon flow-type conductivity measurement โ concentration conversion |
| Sensor input | 1-channel |
| Temperature sensor specifications | Platinum resistance 1000ฮฉ/0ยบC |
| Measuring range | HF concentration: 0 to 10% HCI concentration: 0 to 10% Conductivity: 0 to 1000 mS/cm (without temperature compensation) Temperature: 0 to 100ยบC (The number of decimals displayed can be selected from 1 and 2.) |
| Repeatability | Within ยฑ0.5% of the full scale (in equivalent input) |
| Linearity | Within ยฑ0.5% of the full scale (in equivalent input) |
| Transmission output | No. of outputs: 4 4 to 20 mA DC or 0 to 20 mA DC: I/O insulation type Maximum load resistance: 900ฮฉ Transmission output range: Custom setting possible within measurement range (Negative terminals of each transmission output channels are connected internally thus have the same electric potential.) |
| Contact output | No. of outputs: 5 Alarm contact outputs: R1, R2, R3 and R4 Contact type: Relay contact, SPST Contact rating: 240 V AC, 1A and 30 V DC, 1A (resistance load) Contact function: Selectable between upper/lower limit operation (ON/OFF control), and delay operation Contact action: โข Closed during control action (during error setting) โข Opened during normal status (including power supply off) Error alarm output (RF) Contact type: Relay contact, SPDT Contact rating: 240 V AC, 1A and 30 V DC, 1A (resistance load) Contact function: Selectable from error, trouble warning and maintenance Contact action: When connected between C and NO, the contact is closed when status is normal and opened when error is detected (including power supply off) (Contact status is reversed for connection between C and NC.) (R1, R2, R3 and R4, RF are for common use, respectively.) |
| Contact input | No. of inputs: 1 Contact status: Open collector no-voltage contact a Contact function: Transmission output hold |
| Communications output | RS-485 input/output |
| Calibration function | Conductivity: Based on the specified compensation coefficient for the cell constant (parameter input) Temperature: Calibrated by comparing with the reference thermometer |
| Transmission output Hold function | Selectable from the previous value hold and optional value hold (However, only the previous value hold is available during the maintenance mode. |
| Self-diagnosis function |
|
| Temperature compensation |
|
| Temperature compensation range | HF: 0 to 80ยบC, HCI: 0 to 80ยบC, Conductivity: 0 to 100ยบC |
| Ambient environment | Temperature: -5 to 45ยบC, Relative humidity: 20 to 85% (without condensation) |
| Power supply | 24 V DC ยฑ10%, 10 W (max.) |
| Protective structure | Panel: IP65, Rear case: IP20, Terminal: IP00 (indoor-use panel installation type) |
| Mass | Approx. 550g |
| Conforming standards | CE marking, FCC Part 15 |
| Compatible sensors | FES-510 series (flow-type carbon sensor) |
4-electrode carbon flow-type conductivity sensor
| Model | FES-510-1/4 | FES-510-3/8 | FES-510-1/2 | FES-510-3/4 | FES-510-1 |
|---|---|---|---|---|---|
| Liquid end material | Glass carbon (electrode), PFA (body), Kalrezยฎ (seal) | ||||
| Piping size | 1/4 inch | 3/8 inch | 1/2 inch | 3/4 inch | 1 inch |
| Sample flow rate | 0 to 2 L/min | 0 to 8 L/min | 0 to 10 L/min | 0 to 15 L/min | 0 to 25 L/min |
| Sample pressure range | 0 to 0.5 MP for 5 to 50ยบC, 0 to 0.1 MP for above 100ยบC For 50 to 100ยบC, pressure range is plotted between 0.5 MPa and 0.1 MP as shown on the line graph (refer to diagram on the right) | ||||
| Sample temperature | 5 to 100ยบC (Aqueous solution which boils at 100ยบC) | ||||
| Cable length | 10 m (standard) | ||||