| Models | MEXA-2000SPCS | MEXA-2100SPCS | MEXA-2200SPCS | MEXA-2300SPCS | ||
|---|---|---|---|---|---|---|
| Complying Regulations*1 | GTR No.15, Amendment 6 Euro 6, Euro VI, Stage V (NRMM), China 6, China VI, Bharat Stage VI (LDV/HDV) | - | GTR No.15, Amendment 6 Euro 6, Euro VI, Stage V (NRMM), China 6, China VI, Bharat Stage VI (LDV/HDV) | -*2 | ||
| Measuring principle | Laser scattering condensation particle counting (CPC) | |||||
| Lower particle size limit | Counting efficiency of 23 nm particles: 50 % ยฑ12 %, Counting efficiency of 41 nm particles: >90 % | |||||
| Measuring components and range | Number concentration of solid particles; 0 โ 10000 up to 0 โ 50000 particles/cm3 (after internal dilution) *3 | |||||
| Sample handling temperature | 52 ยฐC or less (Dilute sampling) | Maximum permissive temperature (Direct sampling) 350 ยฐC *4 | 52 ยฐC or less (Dilute sampling) | Maximum permissive temperature (Direct sampling) 350 ยฐC *4 | ||
| Diluted sample temperature | Primary diluter (PND1): 191 ยฐC ยฑ 10 ยฐC Evaporation tube (ET): 350 ยฐC ยฑ 10 ยฐC Secondary diluter (PND2): < 35 ยฐC (at inlet of CPC) | Pre-classifier: 47 ยฐC ยฑ 5 ยฐC Primary diluter (PND1): 191 ยฐC ยฑ 10 ยฐC Evaporation tube (ET): 350 ยฐC ยฑ 10 ยฐC Secondary diluter (PND2): < 35 ยฐC (at inlet of CPC) | Primary diluter (PND1): 191 ยฐC ยฑ 10 ยฐC Evaporation tube (ET): 350 ยฐC ยฑ 10 ยฐC Secondary diluter (PND2): < 35 ยฐC (at inlet of CPC) | Pre-classifier: 47 ยฐC ยฑ 5 ยฐC Primary diluter (PND1): 191 ยฐC ยฑ 10 ยฐC Evaporation tube (ET): 350 ยฐC ยฑ 10 ยฐC Secondary diluter (PND2): < 35 ยฐC (at inlet of CPC) | ||
| Dilution factors in diluters | Primary diluter (PND1): 10 to 200 *3 Secondary diluter (PND2): 15 | Diluter in DSU: 10 Primary diluter (PND1): 10 to 200 *3 Secondary diluter (PND2): 15 | Primary diluter (PND1): 10 to 200 *3 Secondary diluter (PND2): 15 | Diluter in DSU: 10 Primary diluter (PND1): 10 to 200 *3 Secondary diluter (PND2): 15 | ||
| PCRF | 0.95 < fr (30 nm) / fr (100 nm) < 1.3, 0.95 < fr (50 nm) / fr (100 nm) < 1.2 | |||||
| Volatile particle removal efficiency | >99.0%, for C40 (30 nm of particle size, and 10000 particles/cm3 or more) | |||||
| Accuracy of dilution factor | Within ยฑ 10 % of nominal dilution factor setting (for VPR total dilution factor of 150 to 3000, gas based) *5 | |||||
| Operating environment | Without CLU (standard): Ambient temperature: 5 ยฐC to 30 ยฐC, Ambient humidity: 80 % or less as relative humidity With CLU (optional): Ambient temperature: 5 ยฐC to 45 ยฐC, Ambient humidity: 80 % or less as relative humidity | |||||
| Power supply voltage and frequency | 200/220/230/240 V AC (ยฑ10 %, max. 250 V), 50/60Hz (ยฑ1.0 Hz), single phase (to be specified at ordering) | |||||
| Power requirements | Main unit: Max. 2.3 kVA Main unit and all optional units: Max. 4.5 kVA | Main unit: Max. 2.5 kVA Main unit and all optional units: Max. 4.4 kVA | Main unit: Max. 2.4 kVA Main unit and all optional units: Max. 4.3 kVA | Main unit: Max. 2.6 kVA Main unit and all optional units: Max. 4.5 kVA | ||
| Dimensions (excluding any projections) /Mass | ||||||
| Main unit (without transfer tube, control unit and optional units) | 434(W)ร731(D)ร637(H) mm Approx. 115 kg | 434(W)ร845(D)ร637(H) mm Approx. 120 kg | 434(W)ร910(D)ร637(H) mm Approx. 140kg | 434(W)ร910(D)ร637(H) mm Approx. 145 kg | ||
| Optional units | CYU: Approx. 290(W)ร146(D)ร 236(H) mm *6 Approx. 4 kg SRU: Approx. 300(W)ร550(D)ร 450(H) mm Approx. 35 kg CLU: Approx. 570(W)ร850(D)ร1227(H) mm Approx. 80 kg (for CLU and optional cabinet) DFC: Approx. 464(W)ร550(D)ร 320(H) mm Approx. 38 kg LCU: Approx. 350(W)ร690(D)ร 670(H) mm Approx. 35 kg VGU: Approx. 550(W)ร300(D)ร 450(H) mm Approx. 20 kg | |||||
*1: Please contact us for more information on the regulations that the products comply with.
*2: MEXA-2300SPCS can be used in the measurement method according to the regulation. For detailed information, please contact HORIBA.
*3: Dilution factor of the system should be determined so that the particle concentration after dilution fits into the measuring range.
*4: Allowable range of gas temperature at sample probe inlet depends on the sampling condition, because it is limited as the temperature of diluter in DSU (350 ยฐC or less). For detailed information, please contact HORIBA.
*5: Accuracy of dilution factor check with span gas.
*6: The dimensions depend on customers.