ASGU-370P

Automatic Span Gas Unit for Parallel calibration

The multi point calibration unit ASGU-370P is designed for simultaneous (parallel) checking of several gas analyzers. It is primarily installed in laboratories for quality assurance and is also used for the production of gas analyzers.

The calibration unit ASGU-370P will be used for the dynamic and continual manufacture of zero and span gas to perform among other things all ongoing quality control procedures according EN 14211, EN 14212, EN14626 and EN 14625. To produce various span gas the unit uses technologies like dilution, permeation, UV irradiation or Gas Phase Titration.

 

Division: Ambient
Manufacturing Company: HORIBA (Austria) GmbH
  • Multipoint calibration acc. EN 14211, EN 14212, EN14625 and EN 14626
  • High accuracy of span gas in cause of usage of factory calibrated digital thermal mass flow controllers in high quality
  • One 3 HU unit for calibration of several analyzers. Low space requirement.
  • Simultaneous operation in cause of independent layout of channels (modules)
  • Comfortable operation via big, illuminated Touch-Screen-Display or externally from analyzer               
  • Operation via RS232 and Ethernet interface (Bayern Hessen protocol)
  • Delivery scope includes remote software for easy external handling (Ethernet)
  • High flexibility by free definable span points (up to 20)
  • High flexibility by free definable cycles (e.g., Lack of fit)
  • High flexibility by free definable sequences (stringing cycles together)
  • Programmable timer allows time shifted start of calibration process
  • Internal data logger (option) stores raw values from calibration procedure (set values and actual values, status) for reporting with evaluation software (included in scope of delivery)
  • Adapted to customer’s request
  • Modular Layout enables later upgrade with additional span gas modules

Principle

Dilution, Permeation, UV irradiation, Gas Phase Titration, Mixed gas dilution 

Application

Quality control in Laboratory                                                                   

Stage of extension

Up to 6 independent modules possible (parallel operation)

Flow regulation

Digital Mass Flow Controllers

Span gas repeatability

+/- 1%

Span gas linearity

+/- 2% (minimum flow > 20% f.s.)

Permeation System (NO2, SO2, NH3, H2S, BTX)

PID temperature controller with Pt 100 sensor
Temperature: 50 °C
Accuracy ± 0,1 °C
Indication of actual temperature on display

Ozone generator by UV Irradiation

PID temperature controller with Pt 100 sensor
Temperature: 70 °C
Accuracy ± 0,1 °C
Indication of actual temperature on display

Zero Air Supply

External

Operation

Manual (Touch Screen Display)
External from Analyzer (Contact, 24VDC)
External via RS232 (Bayern Hessen Protocol)
External via Ethernet (Remote Software or Bayern Hessen Protocol)

Power

230 VAC, 50Hz 
Power consumption depends on stage of extension

Dimensions

19"(W), 570 mm(D), 3 HU(H)

Mass

Depends on stage of extension

  
Air Quality Monitoring
Air Quality Monitoring
During the period of rapid economic growth, Japan was burdened with various types of pollution such as water contamination and air pollution. The first step to regain the glow of the sky and the purity of water was analysis. HORIBA's measuring instruments play an active part in various fields to determine, for example, whether established regulations are complied with or whether the air and water are purified properly. The beautiful environment is the world's asset. Going forward, HORIBA will continue contributing to preserving the global environment through measurement.

제품 문의

HORIBA제품의 자세한 정보를 원하시면, 아래의 양식에 내용을 입력을 부탁드립니다.

* 는 필수입력항목입니다.

Related products

AP-370 series
AP-370 series

Air Pollution Monitor

AQMS
AQMS

Ambient Air Quality Monitoring System

PX-375
PX-375

PM & 중금속 분석기

Corporate