HC-300F

現場安裝型氟離子濃度計

用於半導體/FPD工廠和玻璃工廠的廢水管理。

氟離子濃度計,覆蓋範圍廣,清洗裝置種類豐富。與氟離子電極1009結合可檢測樣品中的遊離氟離子。非常適合管理半導體/FPD 工廠和玻璃工廠的廢水。

產品分類: Process Water
製造商: HORIBA Advanced Techno Co., Ltd.

Supporting a wide range

With a single unit, a wide range from low to high concentrations can be covered.

Comprehensive range of cleaning devices.

A diverse lineup of cleaning devices are offered.

Converter rating

Product nameFluoride ion meter
ModelHC-300F
Measurable range0 mg/L to 10000 mg/L (Display range: 0 mg/L to 11000 mg/L)
Available range options:
10000, 5000, 2000, 1000, 500, 200, 100, 50, 20.0, 10.0 mg/L
0℃ to 100℃ (Display range: -20℃ to 130℃)
Display
resolution
Concentration0.1 mg/L: 0.0 mg/L to 20.0 mg/L
1 mg/L: 0 mg/L to 200 mg/L
10 mg/L: 0 mg/L to 2000 mg/L
100 mg/L: 0 mg/L to 10000 mg/L
Temperature0.1℃
PerformanceConcentrationRepeatabilityWithin ±7% of the full scale (response for equivalent input)
LinearityWithin ±10% full scale (response for equivalent input)
TemperatureRepeatability±0.3℃ (response for equivalent input)
Linearity±0.3℃ (response for equivalent input)
Transmission
output
Output type4 mA to 20 mA DC, input/output isolated type
Load resistanceMax. 600Ω (case of 24 V DC power supply)*1
LinearityWithin ±0.08 mA (output only)
RepeatabilityWithin ±0.02 mA (output only)
Output rangeConcentration: Selection from preset ranges or free range input within measuring range.
Occasional out for errorHold or burnout to either 3.8 mA or 21 mA
Transmission holdIn the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact inputNumber of input points1
Contact typeNo-voltage a contact for open collector
ConditionsON resistance: 40Ω
Open voltage: 1.2 V DC
Short-circuit current: 21 mA DC max.
Contact functionExternal input for transmission holding.
Temperature
compensation
Applicable temperature elementPlatinum resistor: 1 kΩ (0℃)
Positive relation resistor with temperature: 10 kΩ (25℃)
0℃ to 100℃
Temperature calibration1 point calibration comparing reference thermometer
CalibrationNumber of calibration pointsSelectable from 1, and 2 points
Kinds of standard solutionsThe first point: Calibrated by the standard solution with the density from
50 to 100% of the measuring range.
The second point: Calibrated by the standard solution with the density
from 1 to 20% of the measuring range.
Additional capabilitiesAutomatic detection of calibration failure (asymmetry potential,
sensitivity, or response time)
Calibration history (asymmetry potential, sensitivity, and number of days elapsed after last calibration)
Self-checkCalibration errorAsymmetry potential error, and sensitivity error
Temperature calibration error
Electrode diagnostic errorTemperature sensor short-circuit, temperature sensor disconnection,and out of the temperature measurement range
Meter errorCPU error, ADC error, and memory errors
Operating temperature range-20℃ to 60℃ (without freeze)
Operating humidity rangeRelative humidity: 5% to 90% (without condensation)
Storage temperature-25℃ to 65℃
Power supplyRated voltage24 V DC (operating voltage range: 21 V to 32 V DC)*1
Power consumption0.6 W (Max)
Compatible
standards
CE markingEMC directive: EN61326-1
RoHS Directive: EN50581
EMCImmunity
Industrial
electromagnetic
environment
Electrostatic dischargeIEC61000-4-2
Electromagnetic field at radiated radio frequencyIEC61000-4-3
Electric fast transient/burstIEC61000-4-4
SurgeIEC61000-4-5*2
Conducted interference induced by radio
frequency
IEC61000-4-6
Emission
Class A
Radiated disturbanceCISPR 11 CLASS A
FCC rulesPart15 Class A
StructureInstallationOutdoor installation type
Installation methodMount on 50 A pole or wall
Protection codeIP65IEC60529, JIS C 0920
Case materialAluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittingsSUS304
Material of hoodSUS 304 stainless steel
(coated with epoxy-denatured melamine resin)
Material of windowPolycarbonate
Display elementReflective monochrome LCD
External dimensions180 (W) × 155 (H) × 115 (D) mm (excluding mounting brackets)
MassMain body: Approx. 2.8 kg; cover and mounting brackets: Approx. 1 kg

 

*1: The max. load resistance is available within the range shown in the figure below, depending on power supply voltage:

Relation between power supply voltage and load resistance

*2: When the sensor cable, the transmission cable, or the contact input cable is extended to 30 m or longer, the surge test specified in the EMC directive for CE marking is not applied.

Note

For transmission output, an arrester (spark-over voltage: 400 V) is implemented. However, use the optimal surge absorption element on the connection line in accordance with the ambient environment, equipment installation situation, externally connection equipment, and others.

 

Electrode rating

Product nameFluoride ion electrode
Model1009
Sample water
conditions
pH range4pH to 8pH (at 0.2 mg/L)
4pH to 10pH (at 20 mg/L)
4pH to 12pH (at 2000 mg/L)
Temperature range0℃ to 40℃
Electric conductivity500 μs/cm or more
Flow rate range1 cm/sec to 20 cm/sec
Electrode structureStructureSubmersible type
Response filmLanthanum fluoride
Comparative electrodeAg/AgCl, Internal solution KCL3.3 mol/L, Replenishment type
Liquid junctionCeramic double junction
Temperature
compensation element
Pt1000 Ω (0℃)
Cable length5 m
External dimensions12φ mm × 175 (L) mm (cable and projection excluded)
MassApprox. 200 g
OptionHolderCH-10 (Immersion type holder)
CF-30 (Inline type holder)
Cleaning deviceJCH-12 (Immersion type water-jet cleaning device)
JCH-32 (Circulation type water-jet cleaning device)

 

Note

  • This monitor detects the free fluoride ions in liquids, but it does not detect all fluorides. When complexes are present, a measurement lower than JIS will result.
  • Measurements may vary depending upon the pH level. Use this monitor in close to neutral pH conditions.

留言諮詢

如您有任何疑問,请在此留下詳細需求或問題,我們將竭誠您服務。

* 這些欄位為必填項目。

Related products

Aqualog Datastream
Aqualog Datastream

For Instant Water Quality Reports

HD-960LR
HD-960LR

Dissolved Oxygen Concentration Monitor Series for Semiconductor Manufacturing

ICP Neo
ICP Neo

ICP Software

LAQUA DO210
LAQUA DO210

Handheld Dissolved Oxygen/Temperature Meter

LAQUA DO220
LAQUA DO220

Handheld Dissolved Oxygen/Temperature Meter

LAQUA DS-72
LAQUA DS-72

Benchtop Conductivity/Resistivity/Total Dissolved Solids/Salinity/Temperature Meter

LAQUA EC1100
LAQUA EC1100

Benchtop Water Quality Meters

LAQUA EC210
LAQUA EC210

Handheld Conductivity/Resistivity/Salinity/Total Dissolved Solids/Temperature Meter

LAQUA EC220
LAQUA EC220

Handheld Conductivity/Resistivity/Salinity/Total Dissolved Solids/Temperature Meter

LAQUA F-72
LAQUA F-72

Benchtop pH/ORP/Ion/Temperature Meter

LAQUA F-73
LAQUA F-73

Benchtop pH/ORP/Ion/Temperature Meter

LAQUA F-74
LAQUA F-74

Benchtop pH/ORP/Ion/Conductivity/Resistivity/Total Dissolved Solids/Salinity/Temperature Meter

LAQUA PC1100
LAQUA PC1100

Benchtop Water Quality Meters

LAQUA PD210
LAQUA PD210

Handheld pH/ORP/Dissolved Oxygen/Temperature Meter

LAQUA PD220
LAQUA PD220

Handheld pH/ORP/Dissolved Oxygen/Temperature Meter

LAQUA PH1100
LAQUA PH1100

Benchtop Water Quality Meters

LAQUA PH1200
LAQUA PH1200

Benchtop Water Quality Meters

LAQUA PH1300
LAQUA PH1300

Benchtop Water Quality Meters

LAQUA PH1500
LAQUA PH1500

Benchtop Water Quality Meters

LAQUA PH210
LAQUA PH210

Handheld pH/ORP/Temperature Meter

LAQUA PH220
LAQUA PH220

Handheld pH/ORP/Temperature Meter

LAQUAtwin Ca-11
LAQUAtwin Ca-11

Pocket Water Quality Meters

LAQUAtwin Ca-11C
LAQUAtwin Ca-11C

Pocket Water Quality Meters

LAQUAtwin Ca-11C
LAQUAtwin Ca-11C

Bovine Calcium Meters

LAQUAtwin EC-22
LAQUAtwin EC-22

Pocket Water Quality Meters

LAQUAtwin EC-33
LAQUAtwin EC-33

Pocket Water Quality Meters

LAQUAtwin K-11
LAQUAtwin K-11

Pocket Water Quality Meters

LAQUAtwin Na-11
LAQUAtwin Na-11

Pocket Water Quality Meters

LAQUAtwin NO3-11C/NO3-11S/NO3-11
LAQUAtwin NO3-11C/NO3-11S/NO3-11

Pocket Water Quality Meters

LAQUAtwin pH-33
LAQUAtwin pH-33

Pocket Water Quality Meters

LAQUAtwin Salt-11
LAQUAtwin Salt-11

Pocket Water Quality Meters

514-05
514-05

0.5% NaCl Standard Solution

514-22
514-22

1.41 mS/cm Conductivity Standard Solution

514-23
514-23

12.9 mS/cm Conductivity Standard Solution

514-50
514-50

5.0% NaCl Standard Solution

Corporate