VS70-PDA

VS70-PDA
VS70-PDA optical layout
VS70-PDA spectral coverage
VS70-PDA Quantum Efficiency

高信噪比 PDA 微型光谱仪

VS70-PDA OEM 光电二极管阵列光谱仪是一种高性能紧凑型光纤耦合光谱仪,覆盖了 200 至 1000 nm 的宽光谱范围。 这款适用于工业应用的 VS70 系统使用改进的 VS70 光学系统,带有针对 UV-VIS 优化的 Horiba 的 IV 型像差校正平场全息离子蚀刻凹面光栅和线性光电二极管阵列。v

该系统配备定制的多区域排序过滤器以消除高阶衍射,并采用单光学设计,可实现卓越的成像、高灵敏度、高分辨率和低杂散光性能,适用于强光应用,如发射、吸光度和反射率。

OEM 工业应用

非常适合工业强光应用,例如发射、吸收和反射

• 制药和生命科学

• 半导体

• 液相色谱

• 环境、农业和能源

 

事业部: 科学仪器
产品分类: OEM光谱仪和相机
制造商: HORIBA Scientific
  • 高光谱采集速度(高读出速度 3.5 ms)
  • 从紫外到近红外的宽光谱覆盖
  • 高达 1 Ge– 的巨大 PDA,比任何 CCD 深 100 多倍
  • 高信噪比与超低杂散光相结合
  • 具有平滑变化的光谱响应的高紫外线灵敏度
  • 高光通量
  • 紧凑的尺寸坚固性和稳定性

 

OEM Spectrometer Selection Guide

浏览并查看我们的OEM光谱仪选择指南

General Spectrometer Specifications*

Spectral Coverage

Standard: UV-VIS (200–860 nm) with 477 l/mm grating, VIS (380–750 nm) with 582 l/mm
grating, UV-NIR (200–1050 nm) with double-blazed 365 l/mm grating
On request other spectral ranges available / different mechanical integration might needed

Spectrometer Input

Fiber coupled: FC or SMA; Free space: 12-25-37-50-62-75-100-125-150-200 μm slits, other options
available upon request

Spectral Resolution

2 nm for 365 l/mm grating, 1 nm for 477 l/mm grating, 0.8 nm for 582 l/mm grating
(25 μm slit for all)
Average Spectral
Dispersion
30.9 nm/mm for 365 l/mm grating, 23.7 nm/mm for 477 l/mm grating,
16.3 nm/mm for 582 l/mm grating

Focal Length

70 mm

OptionsSelection of high grade sensors: CMOS, B.I. CCD, PDA
Input port: SMA, FC, free space, custom input

F/#

F/2.3

Stray-light Rejection
Typical (Maximum)
0.01% (0.02%) for UV-VIS configuration with 300 μm tall CCD, measured at 700 nm
(measured with broad bandpass 510 nm filter, 75 μm slit-width)
> 2.4 AU linear range (5% variation) with caffeine 273 nm absorption peak in 10 mm cuvette
and D2 lamp.

Wavelength Accuracy

≤ 0.2 nm (Average Absolute Wavelength Error)

Software**

LabVIEW™ acquisition software for initial evaluation (DLLs provided for software integration)

Detector Options and Specifications*
High sensitivity, high dynamic range linear photdiode array
Detector ModelHamamatsu S10123-1024N with higher UV sensitivity and smoothly varying spectral response
characteristics in UV region and wide photosensitive area
Sensor Format1024 x 1 pixels (512 pixels available on request)
PDA Pixel Size25 x 500 μm (taller 2500 μm version available on request)
PDA Active Area25.6 x 0.5 mm
PDA QESee the plot below
Sensor TemperatureUncooled
Maximum Spectral Rate 
Full Well Capacity100 Me- (1 Ge- optional)
Readout Noise3,900 e- (typical); 5,250 (maximum)
Digitization16-bit
Dynamic Range (FW/RN) 
SNR10,000:1 (typical)
Non-linearity (measured
on each system)
< 0.4% (corrected)
Dark Current200,000 e-/pix/s (typical); 250,000 e-/pix/s (maximum)
CommunicationUSB 2
Environmental ConditionsOperating temperature 15° C to 40° C ambient
Relative humidity <70% (non-condensing)
Storage temperature -25° C to 45° C
Power RequirementsPowered by Samtec connector, 5 VDC

 

* HORIBA Instruments 有持续产品开发的政策,并保留修改部件号、说明和规格的权利,恕不另行通知。 **运行我们的采集软件不需要 LabVIEW™ 许可证。

量子效率曲线

 

机械图纸

下载机械图纸

留言咨询

如您有任何疑问,请在此留下详细需求信息,我们将竭诚为您服务。

* 这些字段为必填项。