MicroMax 384 Microwell-Plate Reader

MicroMax 384 Microwell Plate Reader

The MicroMax 384 is a top-reading microwell fluorescence plate reader accessory that is able to accept plates with up to 384 wells and connects to FluoroMax® or Fluorolog® spectrofluorometers for selectable excitation illumination and emission wavelength detection.

事業セグメント: 科学
製造会社: HORIBA Scientific

MicroMax 384’s high speed allows it to completely scan a96 microwell plate in less than one minute. By moving the microwell plate through stationary optics, the MicroMax 384 insures high sensitivity, excellent accuracy, and high reproducibility.

Light from the excitation and emission monochromators is carried via a fiber-optic bundle to and from the MicroMax 384, thus the user may scan with the main spectrofluorometer and select any excitation and emission wavelength pair for intensity measurements. All control of the MicroMax 384 is automated through FluorEssence™ software.

Custom selection of microwells on the plate is possible through the software.

Optical coupling principle

Software Operation

FluorEssence™ applications specifically for the MicroMax 384.

The MicroMax 384 uses three applications within FluorEssence™:

  • Real Time Display
  • Run Experiment
  • Constant Wavelength Analysis

 

Features

  • Second generation device: more sensitivity ...
  • For plates with up to 384 wells
  • Compatible with FluoroMax® and FluoroLog® spectrofluorometers
  • Fully automatic
  • Controlled by FluorEssence software
Wavelength RangeFull range of spectrofluorometers: 250-850 nm
(other wavelength ranges are available)
Microplate Acceptance384 wells (24 X 16) and 96 well (12 x 8). For non-standard sizes, contact HORIBA Scientific Fluorescence Service
SpeedScans 96 wells in less than 1 minute
Typical Sensitivity25 pM fluorescein with Nanolog®-22
10 nM fluorescein with FluoroMax®
Fiber OpticsBifurcated-slit-end bundles. 1m length.
Randomized fiber distribution in bundle.
Dimensions (cm)17.8 H X 30.5 W X 20.3 D
Electrical Specifications110 VAC at 0.50 A and 60 Hz, single phase
220 VAC at 0.25 A and 50 Hz, single phase
Hardware RequirementsFiber-optic adapter
Communication InterfaceRS-232
Computer RequirementsStandard PC from HORIBA Scientific or equivalent computer with free RS-232 port
Software RequirementsWindows®, version #9 or higher
FluorEssence™, version 2.1 or higher
Instrument CompatibilityNanolog® and Fluorolog®-3
FluoroMax® Plus and previous versions 4, 3 and P
(with appropriate fiber-optic adapter)
CertificationCE

 